Scientific paper A New Voltammetric Method for the Determination of Lercanidipine in Biological Samples Funda Ozturk,1 ibrahim Hudai Tasdemir,2 Deniz Altunoz Erdog an,3 * Nevin Erk4 and Esma Kl ll £3 1 Department of Chemistry, Faculty of Arts and Science, Namik Kemal University, Tekirdag, Turkey 2 Department of Chemistry, Faculty of Arts and Science, Ahi Evran University, Kirsehir, Turkey 3 Department of Chemistry, Faculty of Science, Ankara University, Ankara, Turkey 4 Department of Analytical Chemistry, Faculty of Pharmacy, Ankara University Ankara, Turkey * Corresponding author: E-mail: erdodeniz@gmail.com Tel. No.: +903122121278; Fax No.: +903122232395 Received: 11-05-2011 Abstract Electrochemical behavior and adsorption-diffusion properties of lercanidipine (LCN) on a glassy carbon electrode (GCE) were investigated in a mixture of ethanol-Britton Robinson buffer (BR) using voltammetric methods. From experimental results LCN was found to be reduced irreversibly via a single four-electron process controlled mainly by diffusion with some adsorption contribution at about -0.65 V (vs. Ag/AgCl reference electrode). Therefore, a new, accurate, rapid, selective and simple square-wave cathodic adsorptive stripping voltammetric (SWCAdSV) method could be developed for direct determination of LCN in pharmaceutical preparations, spiked human urine and spiked human serum samples without time-consuming steps prior to drug assay. The peak current of the reduction wave linearly changed with the concentration of LCN in the concentration range between 4.0 x 10-8 molL-1 and 7.6 x 10-6 molL-1 in two different regions where optimum preconcentration potential and optimum preconcentration time were applied as -0.20 V and 90 s, respectively. The limit of detection (LOD) and the limit of quantitation (LOQ) values were found to be 2 x 10-8 molL-1 (0.01 mgL-1) and 6 x 10-8 molL-1 (0.04 mgL-1), respectively. The method was applied to determine the content of LCN in commercial pharmaceutical preparation, spiked human serum and spiked human urine. The method was found to be highly accurate and precise, having a relative standard deviation of less than 10% in all applications. Keywords: Electrochemistry, Cyclic voltammetry, Square-wave adsorptive stripping voltammetry, Lercanidipine, An-tihypertensive drug, Biological samples (spiked). l.introduction Dihydropyridine derivatives are used as calcium channel blockers to control the blood pressure and chronic stable angina pectoris. Lercanidipine, (LCN) 2-[(3,3-dip-henylpropyl)methylamine] -1,1-dimethylethylmethyl 1,4-dihydro-2,6-dimethyl-4-(3-nitrophenyl)-3,5 pyridine di-carboxylic ester (Fig. 1), is a third generation dihydropyri-dine calcium channel antagonist that causes systemic vasodilatation by blocking the influx of calcium ions through L-type calcium channels in cell membranes. LCN is a highly lipophilic antihypertensive drug that is effective and well tolerated in patients with mild to moderate hypertension. It maintains adequate blood pressure control for more than a day with once-daily administration, O' Figure 1. Chemical structure of LCN without invoking unfavorable hemodynamic or sympathetic activity. Early clinical studies suggest that LCN has an-tiatherogenic potential and may also protect against endorgan damage.1' 2 Several analytical techniques such as high-performance liquid chromatography,4-9 liquid chromatography,10-16 thin layer chromatography and densitometry,17 liquid-liquid extraction methods,18 and spectrophotometry methods19-24 have been devised for the determination of LCN in pharmaceutical samples or biological fluids. Some of these reported methods are either not sensitive or tedious and require highly sophisticated instrumentation even though they are sensitive. Additionally, there is a polarographic method3 capable of determining LCN in commercial tablets based on its electroreduction. Reviewing the literature revealed that up to now there is no adsorptive stripping voltammetric method for the assay of LCN in pharmaceutical formulations and biological samples. Voltammetric techniques such as cyclic voltamme-try, differential pulse voltammetry and square-wave vol-tammetry have been proved to be very sensitive for the determination of electroactive molecules including organic and inorganic species, drugs and related molecules in pharmaceutical dosage forms and in biological fluids. These methods are faster, easier to operate and cheaper than most of the other analytical methods; the sensitivity increases upon applying stripping voltammetry. Adsorpti-ve stripping voltammetry has been shown to be an efficient electroanalytical technique for determination of a wide range of electroactive species at nano molar level. Its remarkable sensitivity is attributed to the combination of an effective deposition step with an advanced measurement procedure that generates an extremely favorable signal to noise ratio. It usually involves a simple deposition step and most of the excipients used in pharmaceutical preparations do not interfere in the subsequent determination of the drugs. There are many applications of stripping voltamme-tric methods.25-29 In the present study the investigation of electrochemical reduction behavior of LCN using voltam-metric methods was studied. The current study was also carried out to propose a tentative reaction mechanism. Development of a new validated square-wave cathodic adsorp-tive stripping voltammetric (SWCAdSV) assay method for direct determination of LCN in different samples including pharmaceutical preparations, human serum and human urine was one of the other goals of the present study. 2. Experimental 2. 1. Apparatus All voltammetric measurements such as cyclic vol-tammetry (CV), controlled potential coulometry (CPC), square-wave voltammetry with and without cathodic ad- sorptive stripping mode were carried out using a BAS 100B-instrument electrochemical analyzer. A three-electrode cell system (C3 stand) comprising a glassy carbon electrode with a diameter of 3.0 mm as a working electrode (BAS MF-2012), a platinum wire as an auxiliary electrode (BAS MW-1034) and an Ag/AgCl (in 3.0 molL-1 KCl) reference electrode (BAS MF-2052 RE-5B) was used in all experiments. A three-electrode combination system, consisting of a glassy carbon sieve (approximately 65 cm2 area) as a working electrode, coiled platinum wire as an auxiliary electrode (23 cm) (BAS MW-1033) and Ag/AgCl (in 3.0 molL-1 KCl) reference electrode (BAS MF-2052 RE-5B), was used for bulk analysis. All pH measurements were made with a Thermo Orion Model 720A pH ion meter having an Orion combined glass pH electrode (912600) which had been calibrated with pH 4.13 and pH 8.20 stock buffer solutions before measurement. Double-distilled deionized water was supplied from Human Power I+, Ultra Pure Water System (Produced by ELGA as PURELAB Option-S). All the data were obtained at room temperature (21 ± 2 °C). 2. 2. Reagents and Solutions A standard sample of LCN (99.0%, from Fako) was used to prepare the stock solution of LCN. This solution was prepared by dissolution of a precisely weighed amount of LCN in ethanol yielding a LCN concentration of 2.2 x 10-3 molL-1 (1.35gL-1). Calibration solutions were prepared by diluting the stock solution with the mixture of ethanol and Britton-Robinson buffer (BR) in the volume ratio of 20:80 as solvent-supporting electrolyte mixture, and adjusting the pH value of these solutions (using 0.2 molL-1 NaOH) to desired values between 2.5-5.5 for pH. Ortho-phosphoric acid (Riedel), boric acid (Riedel) and acetic acid (Merck) were used to prepare the BR solution; sodium hydroxide (Merck) for adjustment of the pH of the supporting electrolyte was of analytical reagent grade. Double-distilled deionized water was used in preparations of all the solutions. All LCN solutions were protected from light and used within a day to avoid photochemical decomposition. 2. 3. Preparation and Analysis of Samples Lercadip® (produced by Fako in Turkey) tablets were used in pharmaceutical dosage form containing 10 mg of LCN and some amount of excipients per tablet. Before preparation of tablet solutions, the average mass of a tablet was determined by weighing ten finely powdered tablets. A homogeneous sample, equivalent to one tablet, was weighed and transferred to a 100-ml calibrated flask to which approximately 50 mL of ethanol was added. The contents of the flask were sonicated for 30 min to achieve complete dissolution of LCN. After dissolution the flask was filled to the mark with ethanol and the contents of the flask were centrifuged for 30 min at 1500 rpm. A 10-mL aliquot of the clear supernatant was withdrawn and quantitatively diluted to 100 mL with solvent-supporting electrolyte mixture; this LCN stock solution was kept refrigerated. Calibration solutions were prepared from this stock solution by pipetting volumes from 0.025 to 0.75 ml into the electrochemical cell containing 10.0 mL solvent-supporting electrolyte mixture; after pH adjustment to the desired value LCN was determined. For testing spiked serum and urine samples were used. Actual serum and urine samples obtained from healthy individuals were stored frozen until assay. After gentle thawing, a 1-mL aliquot of serum (or urine) was added to the electrochemical cell containing 9 mL of solvent-supporting electrolyte mixture, followed by spiking with LCN stock solution (0.05-3 ml). After deareation, LCN in the cell was quantified using the least squares linear regression ■0.1 -03 -0.5 -0.7 Potential, V (vs Ag/AgCl) -0.9 Figure 2. Cyclic voltammograms of different LCN solutions in given solvent-supporting electrolyte mixture at pH 4.0 at scan rate of 0.10 Vs-1 (inset: plot of current versus concentration) 2. 4. Voltammetric Procedure In all voltammetric studies, 10 mL of LCN solution in solvent-supporting electrolyte mixture was placed into the electrochemical cell. A newly polished GC electrode was inserted into the system and then cell was deoxygena-ted with purified argon (99.99% purity) for 10 min before the first run and 30s between runs. After deareation and equilibration for 5s, voltammograms were recorded by applying a negative-going scan from 0.00 V to -0.90 V. Before each measurement the working electrode (GC) was polished with alumina and rinsed with ethanol for 30 s in an ultrasonic bath. 3. Results and Discussion 3. 1. Electrochemical Behavior of LCN Electrochemical behavior, diffusion and adsorption properties of LCN were studied using the results of cyclic voltammetry, square-wave voltammetry, and controlled-potential electrolysis. In cyclic voltammetric studies a single reduction peak was observed at a potential of about 0.65 V at pH 4.5 (Fig. 2). There is no peak when a blank BR-ethanol mixture was scanned at the same conditions, and the peak current increases linearly with increasing concentration of LCN (Fig. 2; inset). Therefore, it can be concluded that this cathodic reduction peak is due to a reduction of LCN molecules on the GC electrode. As can be seen from Figure (2), there is no anodic peak at reverse scan, indicating that the electrode reaction is totally irreversible. The influence of the potential scan rate on the catho-dic peak current (ipc) was investigated for 5.0 x 10-5 mol-L-1 LCN in the 0.005-1.000 Vs-1 range. In this range a li- near dependency between cathodic peak current, ip c, and scan rate, u, was found (ip c (|iA) = -30.03 u(Vs-1) -3.94; R2 = 0.9937; Fig. 3, inset A). The linear relationship between peak current and scan rate confirms an adsorption-controlled mechanism. The plot of peak current versus square root of scan rate was also constructed (Fig. 3; inset B) and here too a linear relationship was found (ip(|iA) = -32.47Vu + 3.82; R2 = 0.9967). This relationship indicates that diffusion mechanisms are involved in the electrochemical reaction. Also a plot of the logarithm of the peak current (A) versus the logarithm of the scan rate (Vs-1) was studied. This relationship was found to be linear with a slope of 0.621 (Fig. 3; inset C). The value of the slope is between the theoretical value of 1.0 for adsorbed species and 0.50 for a diffusion-controlled mechanism.25 Some extra studies were carried out to control the adsorption phenomena with regard to the literature30, 31 and as a result we found that i) the ratio of cathodic peak current-to-concentration (ipJC) decreases with increasing concentration, ii) the ratio of the cathodic peak current-to-concentration times scan rate (ip JCv) is nearly constant with increasing scan rate and iii) the ratio of cathodic peak cur-rent-to- concentration timesquare root of scan rate (iJCv12) increases with increasing scan rate. Results of p,c all these experimental investigations suggests that electro-reduction of LCN molecules on the GC electrode is mainly controlled by diffusion with some adsorption contribution. In the present study the effect of potential scan rate on the cathodic peak potential (Ep c) was also investigated. p c The peak potential shifts to more cathodic values with increasing scan rate (Fig. 3). The relationship between the peak potential and the logarithm of the scan rate was found to be expressed by the following equation: Ep (V) = Potential, V (vs Ag/AgCl) Figure 3. Influence of potential scan rate on both cathodic peak current and cathodic peak potential of 5.0 X 10-5 molL-5 LCN (inset: (A) curve of peak current versus scan rate, (B) curve of peak current versus square root of scan rate, (C) curve of logarithm of peak current versus logarithm of scan rate, (D) curve of peak potential versus logarithm of scan rate -0.046logu - 0.65 with R2 = 0.9908 (Fig. 3; inset D). Potential shifting with scan rate supports the irreversibility of the electrochemical reaction under investigation. According to the literature32 the slope of the curve of the peak potential versus the logaritm of the scan rate has a value of 0.0296 V per unit (nac), with ac the cathodic charge transfer coefficient and n the number of electons, and the difference between the peak potential and the half peak potential is 0.0477 V per unit (nac). As can be seen from Figure (3), inset D, the curve of the peak potential versus the logaritm of the scan rate has a slope value of -0.046. By using these experimental results the value of na was calculated to be 0.64. This value was calculated to be 0.67 from the difference of the peak potential and the half peak potential. The value of na was also calculated by using another relation given in the literature for CV studies and given below:26 r- , RT E =k +--- ' (na)F logu (1) This equation shows that the slope of the curve of the peak potential versus the logarithm of the scan rate has a value of RT/(na)F. The value of nac was calculated to be 0.61 from CV studies. In electrochemical studies, pH is one of the variables that commonly and strongly influences the electrochemical behavior of molecules. Therefore, the electrochemical behavior of LCN was studied as a function of pH in the pH range 2.15-5.0. At pH values higher than 4.55 the solubility of LCN decreases dramatically and formation of precipitation begins; more and more ethanol is needed to overcome the solubility problem and therefore effect of higher pH values could not be studied. As can be seen from the square-wave voltammo-grams (SWV) in Fig. 4, the potential of the cathodic peak shifts linearly to more negative values with increasing pH as can be expressed by the following equation: Ep c(V) = -0.051 pH - 0.38 with R2 = 0.9690 (Fig. 4; in inset). The experimental value of the slope of this curve was found to be 51 mV per unit pH in the studied pH range. The slope is very close to the theoretical value of 59 mV per unit pH required under assumption of the 2e-/2H+ or 4e-/4H+ process25, 26 of the electro reduction of LCN. Based on the literature, the following equation was used to find the ratio of the number of protons to the number of electrons (d/n) in the electrode mechanism:29 " nF [ß] nF InM (2) In this equation d and n are the number of protons participating in the reaction and the number of transferred electrons in the electrochemical step, respectively. The ratio of the number of protons to the number of electrons was found to be 0.95 from the slope of the plot of Ep versus pH value. As a result, the same number of electrons -0.2 -0.4 -0,6 Potential, V (vs Ag/AgCl) -0.8 Figure 4. Influence of pH on square wave voltammograms of 1.2 X 10-6 molL-1 LCN (inset: Plot of peak potential versus pH value) and protons are partipicating in the electroreduction of LCN molecules. To find out the surface coverage of the adsorbed molecule, the following relationship proposed for adsorption in cyclic voltammetry was used2' (3) where r is the surface coverage of the adsorbed molecules (in mol cm-2), and others are commonly known parameters with their values.26, 29 The surface coverage of the adsorbed molecule (J) was calculated from the slope of the curve of peak current (A) versus scan rate (Vs-1) according to Eq. 3 and was found to be 2.88 x 10-11 mol cm-2 when 0.005 Vs-1