THE INSTITUTE OF ONCOLOGY, LJUBLJANA, YUGOSLAVIA UNIVERSITY GLINIC OF RADIOBIOLOGY AND RADIOTHERAPY, VIENNA, AUSTRIA INTERSTITIAL WATER HYPERTHERMIA: TEMPERATURE DISTRIBUTION DATA OBTAINED IN ANIMAL EXPERIMENTS COMPARED TO HUMAN APPLICATION MEASUREMENTS Lešničar H, Budihna M, Handl-Zeller L, Schreier K Abstract: An original circulating hot water hyperthermia system tor interstitial treatment has been constructed at the University Glinic tor Radiobiology and Radiotherapy in Vienna, Austria. At the Institute of Oncology in Ljubljana, Yugoslavia, animal experiments as well as a few human treatments were done. Results obtained with 4 X 4 metal needle implants on rabbits and/or pigs thigh showed that a good temperature distribution inside the heated volume could be obtained. The homogeneity of temperature distribution obtained in human applications using plastic tubes was also acceptable. Although somewhat lower temperatures inside the treated volume were measured in human treatments, the minimum temperatures obtained were stili on the cytotoxic level (over 42.5 °C). With some improvements using coolling system tor the tumour surrounding zone the damage otthe healthy tissue inside the treatment volume could be avoided. UDC: 612.57-084 Key words: hyperthermia induced-methods Orig. sci. paper Radiol lugosl 1989; 23: 295-7 lntroduction — In clinical practice with interstitial hyperthermia mostly radiofrequency or microwave applicators are used (1, 2, 3). On the other hand, in some centers the practice of nickel-copper alloy thermoseeds has been developed using conductivity as an energy transfer to the tumor tissue (4). Little is known about the possibility of using circulating hot water as a thermal source in hyperthermia treatment (5, 6). The purpose of this paper is to present interstitial measurements data obtained in animal experiments in comparison to first clinical applications using hot water hyperthermia system. Material and methods — T h e U nit: The interstitial hyperthermia unit KHS 9—W 18 (Oth-mar Handl GmbH, Vienna, Austria) consists of a water reservoir, pressure and suction pump, silicone tubes for water supply into and out of the implant, and of the equipment for temperature and flow control (5). The temperature control system is installed in the thermostate, at the inflow and outflow part of the implant, as well as inside the heated tissue (Fig. 1). The l mplant: For the implant 4X4 metal needles or plastic tubes of diameter 1.6 mm with 10 mm spacing were used. Anim a I E x p eriment s: Hyperthermia was induced in thighs of rabbits (3.5—4 kg), and pigs (10—15 kg). The rabbits were anesthesized by 5 % Thiopental injected intraperitoneal^, and pigs by Halothane inhalation. Temperature measurements were done with 5-point thermocouple inserted through 3 catheters placed along the central (A), paracentral (B) and peripheral (C) lines through the implant. The heated volumes ranged from 27-72 cm3 (Fig. 2). Temperature control thermostat Tin and Temperature control on the implant ^1-5 Temperature control in the tissue Figure 1 — Schematic presentation of KHS 9-W 18 hyperthermia unit with implanted tissue and thermoprobe Received: May 15, 1989 — Accepted: May 16, 1989 295 Lešničar H et al. Interstitial. "C" spacing 10 mm O o -B'V'A". o o 0 o o o 0 ) mm O O m JK- peratures measured in human cancer treatment using plastic tube implants and inflow water temperature of 48 °C (Tables 1, 2). If plastic tube implants were used in animal experiments the inflow water temperature of 48 °C resulted in an increase of measured animal tissue temperatures for approx. 2 °C (Table 3). Fig. 2 — Schematic presentation of an implant. Points A, B, C show the position of the catheters tor the temperature measurement CI i n i c a I A p pi i c a tions: Five patients with inoperable carcinomas of the floor of the mouth and/or mobile tongue were treated with the use of interstitial hot-water hyperthermia unit. Implants of the primary tumor using plastic tubes were done under general anesthesia. A special catheter for 5-point thermocouple measurements was inserted through the implanted area. During hyperthermia, several temperature measurements were done at the points of interest, which geometrically corresponded to the measurement lines of animal experiments. Atthe end of hyperthermia, lr-192 wires were inserted through the same plastic tubes in order to combine hyperthermia with brachyradiotherapy. Standard conditions for hyperthermia treatment: 1) Water temperature in the inflow point was kept constant: a) 45.5 °C and 48°C in animal experiments b) 48 °C in clinical applications 2) Flow rate of the circulating hot water was 2.5 ml/sec. per tube. 3) Heating tirne was 60 min. after the tissue temperature had reached the »steady state« (approx. 1O min. after starting). 4) The values of heat transfer for KHS 9-W 18 calculated from previous testing were 15—20 W Measurement line Tissue temp. — rabbit °C±SD* Tissue temp. -pig °C±SD A 43.6±0.5 43.7 + 0.6 B 43.5±0.5 43.7 + 0.6 C 42.3±0.6 41.8±0.5 * SD — standard deviation Table 1 — Average tissue temperatures in °C as measured in lines A, B and C using metal needle implants on rabbits and pigs thigh with inflow watertemperature of 45.5 °C Measurement Tissue temp. point °C±SD A' 43.4±0.5 B' 42.7±0.3 C' 41.4 + 0.4 Table 2 — Average tissue temperature measured in human treatment — plastic tube implant at inflow water temp. 48 °C Measurement Tissue temp. point °C±SD A 45.3±0.5 B 44.4+0.6 C 43.3±0.5 Results — The average tissue temperatures measured in animal and human implants are shown in Tables 1, 2 and 3. The measurement lines A, B and C, positioned constantly in animal implants, geometrically corresponded to the measurement points A', B' and C' chosen as the points of interest in human hyperthermia treatment (A, A' — central; B, B' — paracentral; C, C' — peripheral). Note that measured tissue temperatures obtained in animal metal-needle implants with an inflow water temperature of 45.5 °C correspond quite well to the tissue tem- Table 3 — Average tissue temperature measured in rabbits thigh — plastic tube implant at inflow water temp. 48 °C Discussion — In previously published data on animal experiments with interstitial water hyperthermia (6) it was pointed out that the use of plastic tube implants instead of steel needle implants resulted in approximately 1—2 °C lower average tissue temperatures. For some technical resons (afterloading with lr-192 wires) we were not able to use steel needle implants. Based on our results, it could be concluded that higher tissue temperatures in normal animal tissue are easier to achieve in comparison with human tumor tissue. The cooling effect of arterial system inside the implanted area in rather small experimental areas (rabbits and piglets) having somewhat lower influence on the heating ability 296 Radiol lugosl 1989; 23: 295-7 Lešničar H et al. Interstitial. as compared to the human tissue could serve as a possible explanation of that. These difficulties could easily be overcome by using somewhat higher inflow water temperatures. Nevertheless, the everage tissue temperature measured in our human hyperthermia treatment was stili on cytotoxic ievel, and homogeneity oi temperature distribution expressed as standard deviation of approximately 0.5 °C is highly acceptable. It appears that formerly designed counter-flow needles and tubes with the possibility of cooling one part of the implanted volume are the most promising novelty in our hyperthermia system (6). This improvement should enable us to heat the tumor zone properly without causing any thermal damage to the surrounding healthy tissue. Povzetek INTERSTICIALNA HIPERTERMIJA Z VODO: PRIMERJAVA TEMPERATURNE PORAZDELITVE NA POSKUSNIH ŽIVALIH Z MERITVAMI PRI ČLOVEKU. Na univerzititni kliniki za radiobiologijo in radioterapijo (Dunaj, Avstrija) so zasnovali in izdelali originalno hipertermično napravo za intersticialno zdravljenje s pomočjo ogrevane vode. Avtorji so nato na,Onkološ-kem inštitutu v Ljubljani skupaj izvedli poskusno študijo na živalskem modelu in nekaj kliničnih aplikacij pri bolnikih. Rezultati implantacije stegna s pomočjo 4 X 4 kovinskih igel pri zajcih in/ali prašičih kažejo, da ta način omogoča dobro razporeditev temperature v ogrevanem vplumnu. Dobljena razporeditev temperature je bila dovolj homogena tudi pri kliničnih aplikacijah, kjer so uporabili plastične cevke. Ceprav so pri le-teh v pregrevanem volumnu namerili nekoliko nižje temperature, so bile tudi dobljene minimalne temperature dovolj visoke, da so dosegale citotoksično raven (preko 42,5 °C). Z nekaj izboljšavami, ki bi omogočale hlajenje področja v okolici tumorja, bi se lahko izognili poškodbam zdravega tkiva znotraj implantiranega volumna. References: 1. Bolmsjö M. Commercially available equipmentfor hyperthermia. In: Watmough DJ and Ross WM ed. Hyperthermia. Glasgow, 1986; 224-42. 2. CossetJM, Dutreix J, Haide C, Gerbaulet A. lnters-titial thermoradiotherapy: a technical and clinical study of 29 implantations performed atthe Institute Gustave-Roussy. Int J Hyperthermia 1985; 1: 3-13. 3. Coughlin TC, Douple EB, Strohbehn JW, et al. lnterstitial hyperthermia in combination with brachy-therapy. Radiology 1983; 148: 285-8. 4. Brezovich 1, Atkinson W. Temperature distribution in tumor models heated by self-regulating nickel-copper alloy thermoseeds. 1984; Med Phys 11 (2): 145-52. 5. Handl-Zeller L, Karcher KH, Schreier K, Handl O. Optimierung interstitieller Hyperthermie Systeme. Strahlentherapie 1986; 1963: 460-3. 6. Lešničar H, Budihna M, Handl-Zeller L, Schreier K. lnterstitial hyperthermia with circulating water — results of animal experiments. In: Karcher K. H. ed. Proceedings of the 4th meeting on progress in ra-diooncology. Vienna, 1988; 263-6. 7. Handl-Zeller L, Karcher KH, Schreier K, Budihna M, Lešničar H. The interstitial Viennese system KHS 9-W 18: homogeneous hyperthermia with simultaneous radiation in deep-seated tumors with integrated heat protection of normal tissue. In: Karcher K. H. ed. Proceedings of the IVth meeting on progress in ra-diooncology. Vienna, 1988; 267-9. Author's adress: Hotimir Lešničar, MD, The Institute of Oncology, Zaloška 2, 61000 Ljubljana Radiol lugosl 1989; 23: 295-7 297