Slov Vet Res 2005; 42 (3/4): 65-70 UDC 619:612.119:616-072.5:636.7 Original Research Paper THE EFFECT OF HIGH ANTICOAGULANT K3-EDTA CONCENTRATION ON COMPLETE BLOOD COUNT AND WHITE BLOOD CELL DIFFERENTIAL COUNTS IN HEALTHY BEAGLE DOGS Alenka Nemec 1*, Marinka Drobnic-Kosorok 2, Janos Butinar 1 Addresses of authors: 1 Clinic for Small Animal Medicine and Surgery, 2 Institute of Physiology, Pharmacology and Toxicology, Veterinary Faculty, Gerbiceva 60, 1000 Ljubljana, Slovenia * Corresponding author, Email: alenka.nemec@vf.uni-lj.si Summary: The sodium or potassium salts of ethylenediaminetetraacetic acid (EDTA), anticoagulants recommended for routine haematological analyses, are known to have deleterious effects on platelets, erythrocytes and leukocyte counts even at recommended concentrations (1,8 mg/mL of blood). Under-filling of tubes containing K3EDTA due to difficulties in sampling that produces higher K3EDTA concentrations than recommended may occur in severely hypotensive, restless, very small or obese dogs and cats. The aim of the present study was to investigate the effect of high anticoagulant K3EDTA concentration on complete blood count (CBC) and white blood cell differential count (WCDC) in healthy beagle dogs. The recommended volume of 3 and 1 mL of blood were put in K3EDTA containing tubes to obtain recommended (1.8 mg/mL) and high (5.4 mg/mL) K3EDTA concentrations (under-filled tubes), respectively. CBC and WCDC were determined with a laser haematology analyser with species-specific software. Presence of platelet clumps was determined by examination of blood smears. The results showed a significant (P<0.05) decrease in hematocrit (HCT) from 0.55 ± 0.043 to 0.48 ± 0.042 (13.4 ± 4.8% decrease) and mean corpuscular volume (MCV) from 74.4 ± 1.1 to 65.4 ± 3.0 fL (11.9 ± 4.2% decrease), and a significant increase in mean corpuscular haemoglobin concentration (MCHC) from 306.5 ± 5.7 to 349.7 ±17.8 g/L (14.0 ± 5.9% increase) in samples with high K3EDTA concentration in comparison with the recommended concentration. No platelet clumping was detected in all examined blood smears. According to the results of our study, we recommend respecting the required blood to anticoagulant ratio, as under-filling of tubes leads to a high final concentration of K3EDTA, which affects values of HCT, MCV and MCHC and thus interpretation of haematological results in dogs. Key words: anticoagulants; edetic acid - pharmacology; hematologic tests; blood cell count; leukocyte count; dogs Introduction Several studies have revealed that various anticoagulants used for collection of blood specimens produce different effects on results of haematologi-cal analyses (1, 2, 3, 4, 5). Ethylenediamino-tetraacetic acid (EDTA) is the anticoagulant recommended for complete blood cell counts (CBC) and white blood cell differential count (WCDC) by the National Committee for Clinical Laboratory Standards (6), principally for its cell preservation properties. Among sodium (Na) and potassium (K) salts of EDTA, the International Council for Received: 8 September, 2005 Accepted for publication: 16 January, 2006 Standardization in Hematology (ICSH) currently recommends the dipotassium salts of EDTA (K2-EDTA) as the anticoagulant for CBC (7). Under optimal conditions (appropriate anticoagulant concentration and analysis within 1-4 h of sample collection), the choice of K2 or K3-EDTA has little effect on the results of CBC and WCDC (8). However, EDTA, even when using recommended EDTA concentrations, is known to have deleterious effects on platelets (9, 10, 11, 12, 13), erythrocytes (1, 13, 14, 15, 16) and leukocyte counts (5, 17, 18, 19) and occasionally dogs and cats have strong aggregations of platelets or white blood cells (WBCs) in EDTA blood tubes (20, 21, 22). The collection of blood in other anticoagulants (e.g., heparin, citrate) sometimes prevents or 66 A. Nemec, M. Drobnic-Kosorok, J. Butinar slows the aggregation allowing more accurate WBC counting (23). The phenomenon of in vitro EDTA-induced platelet (PLT) clumping or agglutination is a well-recognized artefact, commonly reported in human medicine (10, 12, 15, 17, 18, 24, 25). The mechanism of agglutination or adherence (clumping) appears to involve EDTA-dependent platelet agglutinins (specific antibodies) present in the plasma (10, 26, 27, 28, 29, 30). There are few reports on EDTA-induced platelet clumping in veterinary medicine (31, 32, 33). EDTA-induced platelet clumping causes spuriously low PLT counts (pseudothrombocytopenia - PTP), which may lead to erroneous diagnosis, unnecessary and costly additional laboratory examinations, and inappropriate medical and surgical therapy or the unjustified withdrawal of essential medication. PTP may be accompanied by either spuriously high white blood cell (WBC) counts (pseudoleukocytosis) or occasionally spuriously low WBC counts (PLT-WBC adherence -pseudoleukopenia). Pseudoleukocytosis is due to the formation of PLT clumps large enough to mimic white cells on the WBC side of an impedance counters (Coulter S Plus IV/V, Technicon H*6000, Ortho ELT 8), whereas pseudoleukopenia may be due to the "gating out" of large PLT-WBC masses (5, 17, 18). Blood sampling following a recommended blood to EDTA anticoagulant volume ratio is necessary to prevent coagulation. Under-filling is probably the most common phenomenon in clinical laboratories due to problems during blood sampling, particularly when blood is collected from very small, obese, severely hypotensive or restless patients. Under-filling of the tubes containing EDTA results in an excess of EDTA, which can reduce PCV and affect other haematological parameters (1, 2, 15, 23, 34). The aim of the present study was to investigate the influence of high anticoagulant K3EDTA concentration, produced by under-filling of tubes, on CBC and WCDC (white blood cell differential count) in healthy beagle dogs. Material and methods Animals Thirty beagle dogs, eighteen females and twelve males, ranging from 8 months to 3 years, were selected and deemed healthy on the basis of history, results of physical examination and serum biochemical profiles (urea, creatinine, sodium, potassium, total bilirubin, albumin and alanine-aminotransferase - results not shown). Sample collection Venous blood samples were collected from fasted dogs using 5 mL syringes; 3 mL and 1 mL blood samples were put in two separate vacuum tubes containing K3EDTA (5.40 mg) as anticoagulant, prepared for the collection of 3 ml blood. The final concentrations of K3EDTA were: 1.8 mg/mL (recommended concentration) and 5.4 mg/mL (high concentration). Both K3EDTA blood samples for CBC and WCDC determination were stored at room temperature until analysis. Venous blood samples were also collected into plain tubes, which stood for 30 min at 4°C to clot, prior to centrifugation (1200 g for 10 min) and separation of serum. Determination of haematological parameters -CBC and WCDC CBC and WCDC were determined by an automated laser haematology analyser Bayer -Technicon H 1 with species-specific software (H 1 Multi-Species V30 Software). The measurements were performed within 1-4 h after venipuncture (5). CBC included white blood cells (WBC), red blood cells (RBC), haemoglobin concentration (HGB), haematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular haemoglobin (MCH), mean corpuscular haemoglobin concentration (MCHC) and platelets (PLT). WCDC represented a six-part differential (neutrophils (NEUT), lymphocytes (LYMP), monocytes (MONO), eosinophils (EOS), basophils (BASO) and large unstained cells (LUC). The LUC category consists of a heterogeneous population of all large cells that fail to exhibit any peroxidase activity (atypical lymphocytes, immature granulocytes and blasts). Blood smears were made from single drops of blood anticoagulated with EDTA. Staining with May Grunwald-Giemsa solution was followed by microscopic examination of the smears for the presence of platelet clumps. Statistical evaluation Statistical analyses were done by use of the statistical programme SPSS 10.0 for Windows (SPSS - statistical package for social sciences). Means and standard deviations were calculated for all haematological and biochemical parameters. A paired t test was used to determine differ- The effect of high anticoagulant K3-EDTA concentration on complete blood count 67 ences in CBC and WCDC parameters between blood samples with recommended and high K3EDTA concentration. A value of P<0.05 was considered significant. Results The results of haematological analyses showed the influence of high K3EDTA concentration on the following parameters of CBC (HCT, MCV and MCHC), while the parameters of WCDC were not affected (Tables 1 and 2). There was a significant (P<0.05) decrease in HCT of 13.4 ± 4.8% (from 0.55 ± 0.043 to 0.48 ± 0.042) and MCV of 11.9 ± 4.2% (from 74.4 ± 1.1 to 65.4 ± 3.0 fL), and a significant increase in MCHC (from 306.5 ± 5.7 to 349.7 ±17.8 g/L) of 14.0 ± 5.9% in samples with high K3EDTA concentration compared to samples with the recommended final concentration of K3EDTA. Platelet clumping was not detected in any of the blood smears. No significant differences in PLT count (Table 1) were found between samples with recommended and high K3EDTA concentrations. Despite the significant differences in HCT, MCV and MCHC between samples with recommended and high K3EDTA concentrations, the CBC and WCDC parameters (mean values) in both types of samples remained within the normal reference ranges given by the producer of the analyser and data from the literature (35, 36, 37). Discussion Despite appropriate sample collection techniques, under-filling of tubes with blood and platelet clumping remains a common problem in veterinary medicine (31, 38, 39). Under-filling of tubes containing anticoagulant EDTA results in higher concentration of EDTA, which is known to cause spurious interferences, even at the recom- mended concentration, on platelets (5, 9, 10, 11, 12), erythrocytes (1, 13, 14, 15, 16) and leukocyte counts (5, 17, 18, 19). Occasionally, dogs and cats have strong aggregations of platelet or white blood cells (WBCs) in EDTA blood tubes (20, 21, 22, 23). The results of the present study in healthy beagle dogs showed no significant changes in MPV and PLT count between samples with recommended and high K3EDTA concentration. Unlike the present study, significantly lower PLT count has been measured in bulls in samples with high K3EDTA concentrations due to EDTA-induced platelet clumping observed in blood smears (33). In the present study, a significant decrease in HCT and MCV and a significant increase in MCHC was found in under-filled samples with high K3EDTA concentration compared to the recommended K3EDTA concentration. The significant changes of HCT, MCV and MCHC are ascribed to shrinking of erythrocytes in a hypertonic medium. Our results are consistent with the study of Chen et al. (1) who investigated the influence of under-filling of EDTA tubes using different human blood collection volumes. All salts of EDTA are hyperosmolar, which causes water to leave the cells, resulting in cell shrinkage. The higher the concentration of EDTA, the greater the osmotic withdrawal of water from the cells, leading to a reduction in PCV. This discrepancy will lead to a reduction in MCV and increase in MCHC (34). When insufficient blood is collected, such as frequently happens in small, obese, hypotensive and restless dogs, and also in cats, accuracy of HCT, MCV and MCHC is affected if K3EDTA is used as the anticoagulant. Following the results of our study, we recommend respecting the required blood to anticoagulant ratio, as under-filling of tubes leads to a high final concentration of K3EDTA, which affects values of HCT, MCV and MCHC and thus interpretation of haematological results in dogs. Table 1: Influence of K3-EDTA on CBC in dogs WBC RBC HGB HCT* MCV* MCH MCHC* PLT (x109/L) (x1012/L) (g/L) (L/L) (fl) (pg) (g/L) (x109/L) 1.8 mg/mL K3EDTA 9.04± 1.55 7.38± 0.60 168.1± 13.1 0.55 ± 0.043 74.4 ± 1.1 22.8± 0.5 306.5 ± 5.7 299.4± 65.5 5.4 mg/mL K3EDTA 9.48 ±1.71 7.27± 0.55 166.0± 11.9 0.48 ± 0.042 65.4 ± 3.0 22.8± 0.4 349.7± 17.8 271.8± 70.6 *P< 0.05, paired t-test The comparison of CBC values (mean ± SD) in blood samples of 30 healthy dogs with recommended concentration (1.8 mg/mL) and high concentration (5.4 mg/mL) of anticoagulant K3EDTA 68 A. Nemec, M. Drobnic-Kosorok, J. Butinar The comparison of WCDC values (mean ± SD) in blood samples of 30 healthy dogs with recommended concentration (1.8 mg/mL) and high concentration (5.4 mg/mL) of anticoagulant K3EDTA. Statistical comparison using paired t-test did not indicate any significant difference Table 2: Influence of K3-EDTA on CBC in dogs NEUT LYMP MONO EOS BASO LUC % % % % % % 1.8 mg/mL K3EDTA 59.09±7.03 32.50 ± 7.09 2.86 ± 1.19 4.78 ± 3.21 0.18 ± 0.06 0.57 ± 0.24 5.4 mg/mL K3EDTA 58.21±6.83 33.33 ± 6.69 3.13 ± 1.54 4.52 ± 3.23 0.28 ± 0.07 0.52 ± 0.19 References 1. Chen BH, Fong JF, Chiang CH. Effect of different anticoagulant, underfilling of blood sample and storage stability on selected hemogram. Kaohsiung J Med Sci 1999; 15: 87-93. 2. Dubin SE, Piszczek JE, Beard R, Schmukler R. Effects of anticoagulants on packed cell volume measurement. Lab Anim Sci 1976; 26: 586-91. 3. Lewis DC, Meyers KM. Effect of anticoagulant and blood storage time on platelet-bound antibody concentrations in clinically normal dogs. Am J Vet Res 1994; 55: 602-5. 4. Olsen AK, Bladbjerg EM, Jensen AL, Hansen AK. Effect of pre-analytical handling on haematological variables in minipigs. Lab Anim 2001; 35: 147-52. 5. Schrezenmeier H, Muller H, Gunsilius E, Heimpel H. Anticoagulant-induced pseudothrombocytopenia and pseudoleukocytosis. Thromb Haemost 1995; 73: 506-13. 6. National Committee for clinical laboratory standards. Evacuated tubes and additives for blood specimen collection: approved standard. 4th ed. NCCLS Document H1-A4. Villanova, PA: National Committee for Clinical Laboratory Standards, 1996. 7. International council for standardization in haematology. Recommendations of the International council for standardization in haematology for ethyl-enediaminetetraacetic acid anticoagulation of blood for blood cell counting and sizing: expert panel on cytometry. Am J Clin Pathol 1993;100 :371-2. 8. Goossens W, Van Duppen V, Verwilghen RL. K2-or K3-EDTA: the anticoagulant of choice in routine haematology? Clin Lab Haematol 1991;13: 291-5. 9. Bartels PCM, Schoorl M, Lombarts AJPF. Screening for EDTA-dependent deviations in platelet counts and abnormalities in platelet distribution histograms in pseudothrombocytopenia. Scand J Clin Lab Invest 1997; 57: 629-36. 10. Fiorin F, Steffan A, Pradella P, Bizzaro N, Potenza R, De Angelis V. IgG platelet antibodies in EDTA- dependent pseudothrombocytopenia bind to platelet membrane glycoprotein lib. Am J Clin Pathol 1998; 110: 178-83. 11. Lippi U, Schinella M, Modena N, Nicoli M. Unpredictable effects of K3EDTA on mean platelet volume. Am J Clin Pathol 1986; 87 :391-3. 12. Lippi U, Schinella M, Nicoli M, Modena N, Lippi G. EDTA-induced platelet aggregation can be voided by a new anticoagulant also suitable for automated complete blood count. Haematologica 1990; 75: 38-41. 13. Eriksson LE. On the shape of human red blood cells interacting with flat artificial surfaces - the 'glass effect'. Biochim Biophys Acta 1990; 1036: 193-201. 14. Meyer DJ, Harvey JW. Veterinary laboratory medicine: interpretation and diagnosis. Philadelphia: W.B. Saunders Company, 1998: 3-21. 15. Penny RHC, Carlisle CH, Davidson HA, Gray EM. Some observations on the effect of the concentration of ethylenediamine tetra-acetic acid (EDTA) on the packed cell volume of domesticated animals. Br Vet J 1970; 126: 383-9. 16. Pinteric L, Manery JF, Chaudry IH, Madapalli-mattam G. The effect of EDTA, cations, and various buffers on the morphology of erythrocyte membranes: an electron-microscopic study. Blood 1975; 45: 709-23. 17. Lombarts AJPF, De Kieviet W. Recognition and prevention of pseudothrombocytopenia and concomitant pseudoleukocytosis. Am J Clin Pathol 1987; 89: 634-9. 18. Savage RA. Pseudoleukocytosis due to EDTA-induced platelet clumping. Am J Clin Pathol 1983; 81: 317-22. 19. Shimasaki AK, Fujita K, Fujio S, Sakurabayashi I. Pseudoleukocytosis without pseudothrombocytopenia induced by interaction of EDTA and IgG2- M protein. Clin Chim Acta 2000; 299: 119-28. 20. Wilkerson MJ, Shuman W. Alteration in normal canine platelets during storage in EDTA anticoagulated blood. Vet Clin Path 2001; 30:107-113. 21. Zelmanovic D, Hetherington EJ. Automated The effect of high anticoagulant K3-EDTA concentration on complete blood count 69 analysis of feline platelets in whole blood, including platelet count, mean platelet volume, and activation state. Vet Clin Path 1998; 27: 2-9. 22. Norman EJ, Barron RCJ, Nash AS, Clampitt RB. Prevalence of low automated platelet counts in cats: comparison with prevalence of thrombocytopenia based on blood smear estimation. Vet Clin Path 2001; 30: 137-140. 23. Weiss D, Tvedten H. The complete blood count and bone marrow examination: general comments and selected techniques. In: Willard MD, Tvedten H, eds. Small animal clinical diagnosis by laboratory methods. Philadelphia: W.B. Saunders Company, 2004: 14-37. 24. Lombarts AJPF, Zijlstra JJ, Peters RHM, Thomasson CG, Franck PFH. Accurate platelet counting in an insidious case of pseudothrombocytopenia. Clin Chem Lab Med 1999; 37: 1063-66. 25. Vicari A, Banfi G, Bonini PA. EDTA-dependent pseudothrombocytopenia: a 12-month epidemiological study. Scand J Clin Lab Invest 1988; 48: 537-42. 26. Hoyt RH, Durie GM. Pseudothrombocytopenia induced by a monoclonal IgM kappa platelet agglutinin. Am J Hematol 1989; 31: 50-2. 27. Onder O, Weinstein A, Hoyer LW. Pseudothrom-bocytopenia caused by platelet agglutinins that are reactive in blood anticoagulated with chelating agents. Blood 1980; 56: 177-82. 28. Pegels JG, Bruynes ECE, Engelfriet CP, von dem Borne AEG. Pseudothrombocytopenia: an immunologic study on platelet antibodies dependent on ethylene diamine tetra-acetate. Blood 1982; 59: 157-61. 29. Van Vliet HHDM, Kappers-Klunne MC, Abels J. Pseudothrombocytopenia: a cold antibody against platelet glycoprotein GP IIb. Br J Haematol 1986; 62: 501-11. 30. Veenhoven WA, Gerda MB, van der Schans GS, Huiges W, Metting-Scherphuis HE, Halie MR, Nieweg HO. Pseudothrombocytopenia due to agglutinins. Am J Clin Pathol 1979; 72: 1005-8. 31. Grindem C, Breitschwerdt E, Corbett W, Jans HE. Epidemiologic survey of thrombocytopenia in dogs: a report on 987 cases. Vet Clin Pathol 1991; 20: 38-43. 32. Koplitz SL, Scott MA, Cohn LA. Effects of platelet clumping on platelet concentrations measured by use of impedance or buffy coat analysis in dogs. J Am Vet Med Assoc 2001; 219: 1552-6. 33. Nemec A, Snoj T, Cebulj-Kadunc N, Cestnik V. The influence of anticoagulant K3EDTA on CBC (complete blood count) in bulls. Zb Vet Fak Univ Lj 1998; 35: 43-8. 34. Bush BM. Interpretation of laboratory results for small animal clinicians. Oxford: Blackwell Science, 1998: 35-131. 35. Bush BM. Interpretation of laboratory results for small animal clinicians. Oxford: Blackwell Science, 1998: 478-83. 36. Davies DT, Fisher GV. The validation and application of the Technicon H*1 for the complete automated evaluation of laboratory animal haematology. Comp Haematol Int 1991; 1: 91-105. 37. Kaneko JJ, Harvey JW, Bruss ML.. Appendixes. In: Kaneko JJ, Harvey JW, Bruss ML, eds. Clinical biochemistry of domestic animals. 5th ed. San Diego: Academic Press, 1997: 885-905. 38. Tasker C, Cripps P, Mackin A. Estimation of platelet counts of feline blood smears. Vet Clin Pathol 1999; 28: 42-5. 39. Knoll J, Rowell S. Clinical haematology. Vet Clin North Am Small Anim Pract 1996; 26: 981-1002. 70 A. Nemec, M. Drobnic-Kosorok, J. Butinar VPLIV VISOKE KONCENTRACIJE ANTIKOAGULANTA K3-EDTA NA KRVNO SLIKO IN DIFERENCIALNO BELO KRVNO SLIKO PRI ZDRAVIH BIGLIH A. Nemec, M. Drobnič-Košorok, J. Butinar Povzetek: Kalijeve in natrijeve soli EDTA (etilendiaminotetraocetna kislina) se kot antikoagulanti največ uporabljajo v rutinski hematologiji, kljub temu da negativno vplivajo na trombocite, eritrocite in levkocite že pri priporočeni koncentraciji (1,8 mg/mL krvi). Pogosto se zgodi, da zaradi težav pri odvzemu krvi pri zelo majhnih, debelih, hipotenzivnih ali nemirnih psih in mačkah, odvzamemo premajhno količino vzorca krvi v epruvete z antikoagulantom K3-EDTA, kar povzroči višjo konce-tracijo EDTA, kot je priporočena. Namen naše raziskave je bil, ugotoviti vpliv visoke koncentracije K3-EDTA na krvno sliko in diferencialno belo krvno sliko pri psih pasme bigel. V epruvete s K3-EDTA smo odvzeli 3 mL (priporočena količina) in 1 mL (premajna količina) krvi in s tem dobili priporočeno (1.8 mg/mL) in visoko (5.4 mg/mL) koncentracijo K3-EDTA. Krvno sliko in diferencialno belo krvno sliko smo določili z laserskim hematološkim analizatorjem, ki je zasnovan na računalniškem programu, s pomočjo katerega lahko določamo krvno sliko in diferencialno belo krvno sliko ljudem in živalim različnih vrst. Z analizo krvnih razmazov smo ugotavljali morebitno prisotnost skupkov trombocitov. V vzorcih z visoko koncentracijo K3-EDTA smo, v primerjavi z vzorci s priporočeno koncentracijo K3-EDTA, ugotovili statistično značilno (P 0,05) znižanje vrednosti hematokrita z 0.55 ± 0.043 na 0.48 ± 0.042 (13.4 ± 4.8% znižanje) in MCV s 74.4 ± 1.1 na 65.4 ± 3.0 fL (11.9 ± 4.2% znižanje) ter statistično značilno zvišanje MCHC s 306.5 ± 5.7 na 349.7 ± 17.8 g/L (14.0 ± 5.9% zvišanje). Analiza krvnih razmazov ni pokazala prisotnosti skupkov trombocitov v vseh analiziranih vzorcih. Na osnovi rezultatov raziskave priporočamo, da se pri odvzemu krvi upošteva zahtevano razmerje med količino odvzete krvi in količino antikoagulanta, saj visoka končna koncentracija K3-EDTA, ki je posledica premajhne količine odvzetega vzorca krvi, vpliva na vrednost hematokrita, MCV in MCHC ter s tem na interpretacijo hematoloških rezultatov pri psih. Ključne besede: antikoagulanti; edetska kislina - farmakologija; hematološki testi; krvnička, štetje; levkociti, štetje; psi