Technical paper Flavonoid Content Assay: Prevalidation and Application on Plantago L. Species Renata Jurisic Grubesica, Jadranka Vukovica*, Dario Kremerb, Sanda Vladimir-Knezevicc a Department of Analytics and Control of Medicines, Faculty of Pharmacy and Biochemistry, University of Zagreb, A. Kovacica 1, HR-10000 Zagreb, Croatia b Department of Pharmaceutical Botany, Faculty of Pharmacy and Biochemistry, University of Zagreb, A. Kovacica 1, HR-10000 Zagreb, Croatia c Department of Pharmacognosy, Faculty of Pharmacy and Biochemistry, University of Zagreb, Marulicev trg 20, HR-10000 Zagreb, Croatia * Corresponding author: E-mail: jadranka@pharma.hr; Tel/Fax: +555 1 49 200 89 Received: 03-07-2006 Abstract This work is aimed to prevalidate and apply UV/Vis spectrophotometric procedure for flavonoid determination in herbal material using AlCl3 as a reagent. Fast and simple, full prevalidation for quality control and standardization of analytical procedure, based on mathematical/statistical testing coupled with system of diagnosis was used to evaluate and demonstrate the reliability of method for flavonoid determination with AlCl3 (F-Al procedure). Favourable prevalidation characteristics verified this procedure as a valuable tool in flavonoid analysis, so it was successfully applied for determination of flavonoids in leaves, stems and flowers of Plantago L. species growing in Croatia. The results showed the various flavonoid contents between different plant parts (leaves: up to 0.13%; stems: up to 0.07% and flowers: up to 0.07%). The results of flavonoid determination were statistically evaluated by using Principal Component Analysis (PCA) and Student's t-test as a contribution to investigation of different taxa of genus Plantago L. Keywords: Prevalidation strategy, flavonoids, aluminium chloride, Plantago L. 1. Introduction Unavoidable part of fundamental and applied investigations in pharmaceutical analysis is analytical procedure with good performance characteristics. The ever-increasing volume of analytical literature concerning quality control requires unambiguously evaluation of the advantages and disadvantages of an analytical procedure.1-9 For this reason, complete prevalidation,10 as an informative screening method, should be useful for preliminary evaluation of an analytical process with regard to reasonable need for validation and for systematically obtaining other valuable data. The aim of prevalidation proposal based on peculiar approaches is to diagnose the quality of an analytical procedure and to decide whether a method in question is capable of producing accurate and reliable data. Investigation of dependent and independent variables, as components of analytical system, particularly relation- ship between them gives insight into the data quality and method's metrological characteristics. Prevalidation is essential to test data validity, e.g. when validate (official) procedure might not exist, when insufficient time would be available for a full validation process, when an analytical method is adopted from some other source or in crisis situations. The efficiency of prevalidation procedure is given by characteristic data such as constants of calibration and analytical evaluation function, standard deviation of procedure, limit of quantitation, and other metrological characteristics. One part of the present study included application of prevalidation strategy to obtain metrological characteristics and verify spectrophotometric procedure for determination of flavonoids with AlCl3 (F-Al procedure). Flavonoids are polyphenolic compounds that occur ubiquitously in plant tissues in relatively high concentrations as sugar conjugates.11 They occur mostly in O-glycosidic form with a number of sugars such as glucose, galactose, rhamnose, arabinose, xylose and rutinose. The flavonoid functions in plants are believed to be as protective agents against UV radiation and also against microorganisms.12 Flavonoids are of particular importance in the human diet as there is evidence that they act as antioxidants,1315 antiviral agents16 and epidemiological studies have indicated that their consumption is associated with a reduced risk of cancer17-20 and cardiovascular disease.21 Another part of these investigations comprehended application of prevalidated F-Al procedure for flavonoid analysis in Plantago L. species growing in Croatia. The genus Plantago comprises 265 species and has cosmopolitan distribution.22 Medicinally, Plantago species are astringents, demulcents, emollients, expectorants, diuretics, antibacterials and antivirals.23 Phytochemical investigations of Plantago species revealed the presence of irido-ids, flavonoids, tannins, triterpenes, saponins, and ste-rols.23-26 The spectrophotometric assay based on aluminium chloride complex formation is one of the most commonly used analytical procedures applied to flavonoid content determination in various plants.27-30 This procedure includes hydrolysis of glycosides, extraction of total flavo-noid aglycones with ethyl acetate and complex formation with AlCl3.31-33 As there are no literature data concerning spectrophotometric determination of total flavonoids in Plantago species, application of F-Al procedure was used to provide new information regarding phytochemical characterization of these plant species. 2. Experimental Experimental comprises protocols for prevalidation of F-Al procedure, then extraction and determination of flavonoids in Plantago species using prevalidated F-Al procedure, as well as multivariate analysis of the obtained results. 2. 1. Apparatus UV/Vis spectrophotometer Agilent 8453 (Agilent, Germany) with PC-HP 845x UV-Visible System (Agilent, Germany) and 1 cm quartz cells were used for all ab-sorbance measurements. 2. 2. Reagents and Solutions Pro analysi chemicals, as well as double distilled water were used throughout the work. Acetone (Kemika, Croatia), 25% hydrochloric acid (Kemika) and hexameth-ylenetetramine (Kemika) were used for the hydrolysis of flavonoide glycoside and plant material extraction. Ethyl acetate (Kemika) was used for aglycone extraction. Aluminium chloride hexahydrate (Kemika) was used as complexing agent. Sodium citrate (Kemika), methanol (Kemika) and acetic acid (Kemika) were used for sample preparation. Filtration of prepared sample solutions was performed by using 0.20 pm Minisart-plus membrane filter (Sartorius AG, Germany). 2. 3. Analytical Standards for Prevalidation Analyte stock standard solution was prepared by exact weighing of 0.01 g quercetin (Roth, Germany), dissolving in 5% solution of acetic acid in methanol and diluting to 100.0 mL with the same solvent. In adequate volume of standard stock solution of quercetin (2.40, 1.94, 1.45, 0.97, 0.48, and 0.24 mL, corresponding to 0.240, 0.192, 0.144, 0.096, 0.048, and 0.024 mg of quercetin, respectively) 0.5 mL of 0.5% sodium-citrate and 2 mL of aluminium chloride was added. Each solution was made up in 25 mL volumetric flask with 5% acetic acid in methanol. After 45 min, the absorbance at 425 nm of the solution was measured. Corresponding compensation solution was prepared and measured identically, but without aluminium chloride. Blank solution was prepared and measured identically, but without analyte. 2. 4. Plant Material Randomly selected samples of wild growing plants of Plantago L. species were collected in the western part of Croatia in June 2003: P. altissima L. in the Mirna River Basin (the north-west of peninsula Istria) at altitude of 20 m; P. coronopus L., P. lagopus L., and P. maritima L. near Medulin (small town in the south of Istria) at altitude of 29 m; P. holosteum. subsp. depauperata Pilger between Vodnjan and Bale (villages in the south of Istria) at altitude of 125 m; P. holosteum subsp. scopulorum (Degen) Horvatic on the islands of Cres and Losinj, near small town Osor, at altitude of 10 m; P. argentea Chaix and P. holosteum Scop. subsp. holosteum on the pass Gornje Jelenje (continental part of the West Croatia) at altitude of 880 m. All plant samples were identified at the Department of Pharmaceutical Botany, Faculty of Pharmacy and Biochemistry, University of Zagreb, Croatia. Voucher specimens (No. 0071-0078) are deposited in the Herbarium of the Department of Pharmacognosy (Faculty of Pharmacy and Biochemistry, University of Zagreb, Croatia). Air-dried samples of leaves, stems, and flowers were phytochemically investigated. 2. 5. Execution of Prevalidation Standardized measurements were based on a set of 24 blocks of data (6 sets of 4 experiments each) to relate measured values to blank values. Samples were measured in standard working range of one power of ten, alternately in the following group sequence: 1, 6, 2, 5, 3, and 4 (Table 1). Table 1: Strategy of prevalidation measurements on standardized basis (N = 24 measurements). Measurements as a process of obtaining results Type of measurements Number of analytical groups Group volume Total number of measurements Analyte amount (quercetin) Analyte working range Sequence of groups measurements Measure (net signal) Gross signal Blank signal Blank measurements (B), gross measurements (y) J = 6, j = 1, 2, ..., 6 I = IV, i = I, ..., IV N = J x I = 24, n = 1, 2, ..., 24 x (mg) 1.0 x, = x1 = xy = 0.240 mg, upper level of analyte, 0.8 x, = x2 = 0.192 mg, 0.6 x, = x3 = 0.144 mg, 0.4 x, = x4 = 0.096 mg, 0.2 x, = x5= 0.048 mg, 0.1 xu = x6 = xL = 0.024 mg, lower level of analyte 1,6, 2, 5, 3, 4 5 = y - B y B Starting data used for mathematical/statistical evaluation were absorbances obtained by measurement of blanks, standards and compensation solutions. Because of possible influence of quercetin on absorbance of quercetin-AlCl3 complex, prior to mathematical/statistical testing gross signal is corrected with absorbance obtained in measurement of corresponding compensation solution. Compensation solution contained the same quantity of quercetin and other components as corresponding analyte solution except aluminium chloride. Systematic prevali-dation strategy and all algorithms together with system of diagnosis were quoted gradually in the paper.10 Mathematical/statistical testing comprised descriptive and prognostic statistics. Arithmetic means, standard and relative standard deviations were used for characterization of all analytical groups (1-6). Prognostic statistics included: checking of groups 1 and 6, testing of data homogeneity, estimation of calibration and analytical evaluation function, outlier recognition and estimation of limiting values. The application of expert system to evaluation of spectrophotometric procedure for determination of quercetin with aluminium chloride was presented in section Results and Discussion in Tables 2-9. Test statistic values were referred to as requirements R throughout the paper. 2. 6. Extraction and Determination of Total Flavonoids The content of total flavonoids (quercetin type) in Plantago species was determined by F-Al procedure (by using method according to Christ and Müller).31 Powdered plant material (0.20 g of each leaves, stems and flowers) was extracted with 20 mL of acetone, 2 mL of 25% HCl and 1 mL of 0.5% hexamethylenetetramine (boiling water bath, 30 min). Each extract was filtered and extraction of the same herbal material was repeated three times with 20 mL of acetone (boiling water bath, 10 min). After cooling and filtration each extract was made up to 100.0 mL with acetone (basic sample solution, BSS). 20 mL of BSS was mixed with 20 mL of water and then ex- tracted with ethyl acetate (first with 15 mL and then three times with 10 mL). Ethyl acetate extracts were rinsed two times with water then filtered and made up to 50.0 mL with ethyl acetate (Solution 1, S1). In 10 mL of S1 0.5 mL of 0.5% solution of sodium citrate and 2 mL of AlCl3 (2 g of AlCl3 in 100 mL of 5% acetic acid in methanol) were added and then made up to 25.0 mL with 5% methanolic solution of acetic acid (sample solution, SS). The same procedure was performed with blank sample solution but without AlCl3. After 45 minutes, yellow solutions were filtered and absorbance at 425 nm was measured. The content of total flavonoids was evaluated upon three independent analyses. The yield was calculated as quercetin toward following expression % = A x 0.772 / b, where A is absorbance and b represents mass of dry herbal material in grams. 2. 7. Statistical Analysis The results of flavonoid analysis were evaluated using Student's i-test and multivariate analysis.34 36 The Principal Component Analysis (PCA) calculation was based on the correlation matrix between the values of the characteristics, which means that the contribution of each variable was independent of the range of its values.37-39 The statistical analysis of the results of flavonoids determination was performed using software Statistica 6.0. 3. Results and Discussion 3. 1. Analysis of Prevalidation Results The analytical signal y, proportional to the absolute mass of the quercetin present, as well as signals obtained from compensation and blank solutions were transformed into the corresponding absorbance values which were used for calculation. Starting data were: mass of quercetin, x, within the working range from 0.024 to 0.240 mg, absorbances obtained in measurements of the blank (B), the sample (y) as well as calculated neto absorban-ce (5). 3. 2. Characterization of Groups 1 to 6 Standardized measurements and calculated values of F-Al procedure were given in Table 2. Standard deviation or relative standard deviation values obtained for all kind of absorbances in each experimental group were used as a measure of precision.8' 40 Reasonable precision in accordance with prevalidation criteria41 was attained for the absorbances obtained in measurements of the sample (s from ± 0.72 to ± 6.57) and for corrected absorbances (srJ, from ± 0.65 to 7.22). Reasonably, the highest values of relative standard deviation' as a reference to reduced precision, were obtained in the group with the smallest analyte content, x6. It is obvious from the results that fluctuations obtained in the measurement of blank samples effect the lower level of precision (srB from ± 14.44 to ± 30.59) and influence on the quality of results could be expected. In the case of great fluctuations of blanks, the influence of blanks could be neglected only if they are small enough in relation to gross values. Table 2: Standard measurements for F-Al procedure. Therefore, additional checking is necessary to conclude about this type of influences (see R7 and R8, Table 4). 3. 3. Checking of Limiting Groups 1 and 6 The preliminary check of working range-limiting groups represented quality control of measurement in a group with the smallest mass of analyte and enables unambiguous distinction between gross and blank signal at x6 (R1, Table 3). Applicability of this requirement was also extended to the recognition of influence of blank values dispersion on the standard deviation of the procedure (sM) through heuristic requirement R2 (Table 3). Preliminary information obtained in this evaluation showed that determination limit is expected below x6 (R3, Table 3). For the standard measurement, requirement that sr values for both gross and corrected measurements at x0 and xL lie below ± 2.5 and ± 25%, respectively was satisfied (R3, Table 3). For the F-Al system under study gross signals could be clearly distinguished from blank signals at x6, although high values of standard deviation of blanks were obtained. Furthermore, total sr value for blanks is 27.8% which corresponds to prevalidation acceptance criteria (total sr < ± 50%).10 Gro- Sam- xa, up ple No. (mg) (j) (i) B B/ sB / srB,% y y/s / s ,% j y ry7 S S / ss / srS,% Ab A/Sa / SrA,% 1 I 0.240 0.0043 0.0043/± 0.0009/ 0.6999 0.6942/± 0.0050/ 0.6956 0.6899/± 0.0049/ 2.8983 2.8745/± 0.0187/ II 0.0056 ± 21.40 0.6969 ± 0.72 0.6913 ± 0.65 2.8804 ± 0.65 III 0.0036 0.6900 0.6864 2.8600 IV 0.0037 0.6899 0.6862 2.8592 6 I 0.024 0.0032 0.0037/± 0.0005 0.0697 0.0643/± 0.0042/ 0.0665 0.0605/± 0.0044/ 2.7708 2.5219/± 0.1820/ II 0.0034 ± 14.44 0.0594 ± 6.57 0.0560 ± 7.22 2.3333 ± 7.22 III 0.0039 0.0637 0.0598 2.4917 IV 0.0044 0.0642 0.0598 2.4917 2 I 0.192 0.0055 0.0041/± 0.0013/ 0.5604 0.5566/± 0.0047/ 0.5549 0.5525/± 0.0042/ 2.8901 2.8777/± 0.0022/ II 0.0045 ± 30.59 0.5521 ± 0.84 0.5476 ± 0.77 2.8521 ± 0.77 III 0.0039 0.5609 0.5570 2.9010 IV 0.0025 0.5531 0.5506 2.8677 5 I 0.048 0.0030 0.0028/± 0.0006/ 0.1175 0.1230/± 0.0062/ 0.1145 0.1202/± 0.0062/ 2.3854 2.5037/± 0.1299/ II 0.0020 ± 20.10 0.1220 ± 5.06 0.1200 ± 5.19 2.5000 ± 5.19 III 0.0033 0.1206 0.1173 2.4438 IV 0.0030 0.1319 0.1289 2.6854 3 I 0.144 0.0043 0.0051/± 0.0011/ 0.4060 0.4097/± 0.0074/ 0.4017 0.4047/± 0.0076/ 2.7896 2.8101/± 0.0530/ II 0.0058 ± 21.43 0.4186 ± 1.81 0.4128 ± 1.88 2.8667 ± 1.88 III 0.0062 0.4017 0.3955 2.7465 IV 0.0040 0.4126 0.4086 2.8375 4 I 0.096 0.0046 0.0035/± 0.0010/ 0.2807 0.2766/± 0.0071/ 0.2761 0.2731/± 0.0063/ 2.8760 2.8451/± 0.0659/ II 0.0040 ± 28.19 0.2812 ± 2.57 0.2772 ± 2.31 2.8875 ± 2.31 III 0.0024 0.2661 0.2637 2.7469 IV 0.0030 0.2785 0.2755 2.8698 6 groups mean s (sr, %) ± 0.0009 (± 23.31) ± 0.0059 (± 3.65) ± 0.0057 (± 3.85) ± 0.0983 (± 3.85) a Mass of quercetin b Measure of particular sensitivity, An = SJxn Table 3: Checking of limiting groups 1 and 6. Requi- Result rement No. Diagnosis R1 AC = 17.25 Significant influence of blank dispersions on sM is not expected R2 R = 162.48% srB1 = ± 21.40% srB6 = ± 14.44% R3 sy6 = ± 6.57% Determination limit is expected below x6 srS6 = ± 7.22% Syt = ± 0.72% srS1 = ± 0.65% LDG = 0.0139 mg LDG = 0.0028 sL = ± 25.25% R4 R = 12.71 Very good resolution of signals R5 R = 3.85L inear calibration function is not expected values (R6, Table 4). The influence of blank values is almost negligible because they are small enough in relation to information obtained at the upper analyte level (R7, Table 4) and total standard deviation of blank values (srBN) was not exceeded ± 50% (R8, Table 4). Since requirements R6, and/or R7, and R8 were satisfied, influence of blank values on results could be excluded and each y value could be corrected with grand blank mean (BN) in F-Al procedure. Barttlet test, applied to s and sr values for B, y, S, A values (R9, Table 4), as well as to the values of the apparent mass of analyte, x (Table 8) provides an insight into the data structure and enables quick recognition of the source of error. For the F-Al procedure under study, Barttlet test was pointed to high data homogeneity of standard and relative standard deviations for majority of values. Lower level of homogeneity was attained for relative standard deviations of gross signals which influence lower homogeneity of neto signals (R9, Table 4). Additional checking of quality of signal resolution for the F-Al procedure showed that gross and blank signals were very good distinguished (R4, Table 3). The preliminary linearity check was applied to A values (particular sensitivity values, A = S/x) for limiting groups 1 and 6. Although the obtained value (R5, Table 3) was very close to tabulated value t (3.707) for this requirement, linear calibration function is not expected. Since only two limiting groups were included in this requirement, systematic and deep evaluation is unavoidable. 3. 4. Testing of Data Homogeneity Analysis of variance applied to the 6 groups of blank values in F-Al procedure indicated homogeneity of blank Table 4: Testing of data homogeneity. Requirement No. Result Diagnosis R6 s2Bb = 2.34 x 10-6 s2Bw = 8.65 x 10-7 R = 2.71 Homogeneous blank values R7 Bn should be < 0.0035 Influence of blank value is not negligible Bn = 0.0039 R8 SrBN = ± 27.77 sBN = 1.09 x 10-3 R9 R(SB) = 2.85 s.h.a R(srB) = 1.78 s.h. R(sy) = 1.41 s.h. R(sry) = 17.12 a.hb R(sS) = 1.63 s.h. R(sJ = 19.60 a.h. R(Sa) = 17.42 a.h. R(sJ = 19.60 a.h. as.h. - strongly homogenous b a.h. ■ - almost homogenous 3. 5. Relation Between Signal and Concentration The characteristic data evaluated by preliminary inspection of the relationship between signal values and content of analyte (method of the least squares) were: determination coefficient (r2), slope of a line (b), intercept of a line (a), errors in the slope (sb), and errors in intercept (sa) (R10, Table 5). The position of the grand mean of sig- Table 5: Quality of relationship analyte amount - analytical signal. Analyte-signal relationship Requi- Result rement No. Diagnosis R10 r = 0.99957 b = 2.9354 a = - 0.0139 s = ± 0.00104 y sb= ± 0.08630 s = ± 0.00041 a centroid = (0.4133, 0.3503) R11 R = 159.54 R12 ± Cb = 2.9354 ± 0.08630 ± C = -0.0139 ± 0.00041 a Significant correlation t-testing for reality of calibration constants R13 V = 2.85 RV = 203.24 sV = ± 0.01405 s m = ± 0.0100 S = 2.85x Ideal calibration function t-testing for reality of analytical evaluation constants R14 V = 0.35 RV = 203.24 sV = ± 0.00172 sM = ± 0.0035 x = 0.35S Ideal analytical evaluation function nal values SN, and the grand mean of mass of analyte jcN, is known as the centroid of all the points. Significance of determination coefficient using statistical t-test showed that for the F-Al procedure significant correlation does exist (R11, Table 5). Errors in the slope and intercept of the regression line were used to estimate confidence limits for the slope and intercept (R12, Table 5). Since the method of the least squares a priori assumed a linear relationship between analytical signal and analyte content, complete and deep evaluation of calibration function using standardized mathematical/statistical procedure was performed.10 The characteristic data evaluated from this procedure were the constants of the calibration and analytical evaluation function, the mean errors of the constants and the standard deviation, sM, of the analytical procedure in the given working range (R13 and R14, Table 5). For the system under study, both ideal calibration and analytical evaluation functions were found. From the final calibration and analytical evaluation function it was possible to evaluate apparent signal values, S, and apparent masses of analyte, x, respectively. Furthermore, analytical functions were used for recognition of outliers and evaluation of analyte limiting values. 3. 6. Outlier Recognition Testing for the outlier was done by comparison of IS* I and I x* I values with the t-values of confidence interTable 8: Data structure for F-Al procedure. vals for P = 95 and 99% confidence level.10 According to prevalidation acceptance criteria, one outlying value is tolerable within the 24-data population. Since one outlying value is observed in F-Al procedure, there is no objection on the homogeneity of the data material (R15, Table 6). Table 6: Test for outliers. Requirement No. Result Diagnosis R15 1 S*31 > 2.069 One outlying value, |x*3| > 2.069 No objection on data material One outlying value, No objection on data material 3. 7. Estimation of Limiting Values According to Gottschalk approach,42 calculation of Ldg was based on sM value of analytical evaluation function and for the F-Al procedure gives the value of LDG = 0.0139 mg of quercetin. This calculated value being bellow the respective x6 level confirmed the correctness of preliminary test R3 (Table 3). According to3, 41 limit of detection and related quantities comprise the slope of the analytical calibration function (sensitivity), V, the total standard deviation of blank values, sBN, and k stands for suggested numerical factor of 3.3 and 10 for the limit of detection, LD, and limit of quantitation, LQ, respectively. All estimated limiting values were significantly lower then the mass of quercetin at lower analyte level, x6 (R16, Table 7). j I S S AS S* x x x sx % Ax Ax/x x 100, % Ax Ax /x x 100, % x* 1 I 0.6960 0.6851 - 0.0105 1.0499 0.240 0.2436 0.2416 ± 0.002 ± 0.65 + 0.0036 + 1.48 + 0.0016 + 0.64 1.0115 II 0.6930 - 0.0062 0.6208 0.2420 + 0.0020 + 0.85 0.5826 III 0.6861 - 0.0013 0.1319 0.2403 + 0.0003 + 0.14 0.0938 IV 0.6860 - 0.0011 0.1120 0.2403 + 0.0003 + 0.11 0.0739 6 I 0.0665 0.0685 + 0.0020 0.2003 0.024 0.0233 0.0212 ± 0.002 ± 7.22 - 0.0007 - 2.98 - 0.0028 - 11.70 0.2041 II 0.0560 + 0.0125 1.2480 0.0196 - 0.0044 - 18.30 1.2515 III 0.0560 + 0.0087 0.8689 0.0209 - 0.0031 - 12.76 0.8724 IV 0.0598 + 0.0087 0.8689 0.0209 - 0.0031 - 12.76 0.8724 2 I 0.5549 0.5481 - 0.0068 0.6822 0.192 0.1943 0.1935 ± 0.001 ± 0.77 + 0.0023 + 1.19 + 0.0015 + 0.76 0.6516 II 0.5476 + 0.0005 0.0461 0.1917 - 0.0003 - 0.14 0.0766 III 0.5570 - 0.0089 0.8918 0.1950 + 0.0030 + 1.57 0.8611 IV 0.5506 - 0.0025 0.2532 0.1928 + 0.0008 + 0.41 0.2227 5 I 0.1145 0.1370 + 0.0225 2.2466 0.048 0.0401 0.0421 ± 0.002 ± 5.19 - 0.0079 - 16.48 - 0.0059 - 12.34 2.2536 II 0.1200 + 0.0170 1.6978 0.0420 - 0.0060 - 12.47 1.7049 III 0.1173 + 0.0197 1.9672 0.0411 - 0.0069 - 14.44 1.9743 IV 0.1289 + 0.0081 0.8098 0.0451 - 0.0029 - 5.98 0.8172 3 I 0.4017 0.4115 + 0.0094 0.9326 0.144 0.1407 0.1417 ± 0.003 ± 1.88 - 0.0034 - 2.33 - 0.0023 - 1.61 0.9552 II 0.4128 + 0.0018 0.1749 0.1445 + 0.0005 + 0.37 0.1521 III 0.3955 + 0.0156 1.5512 0.1385 - 0.0055 - 3.84 1.5736 IV 0.4086 + 0.0025 0.2441 0.1431 - 0.0009 - 0.65 0.2669 4 I 0.2761 0.2740 - 0.0021 0.2064 0.096 0.0967 0.0956 ± 0.002 ± 2.31 + 0.0007 + 0.70 - 0.0004 - 0.39 0.1911 II 0.2772 - 0.0032 0.3162 0.0971 + 0.0011 + 1.10 0.3009 III 0.2637 + 0.0103 1.0308 0.0923 - 0.0037 - 3.82 1.0458 IV 0.2755 - 0.0015 0.1465 0.0965 + 0.0005 + 0.48 0.1313 Barttlet test for x : R(s) = ± 1.63, s.h.; R(Sr) = ± 19.60, a.h., Six groups mean for x: s = ± 0.0020; sr = ± 3.85% Table 7: Estimation of limiting values. Requirement No. Result Diagnosis R16 Ideal calibration function S = 2.85x SD = 0.0072 Ld = 0.0011 mg _Lq = 0.0038 mg SD is expected below S6 Lq is expected below x6 Analysis of variance, the Barttlet test, reality of linear analytical evaluation function and agreement of actual, x, and appropriate, x values gave information on quality of the analytical procedure. With defined analytical evaluation function, it was possible to conclude on accuracy as a total error of analytical procedure using random deviations as well as absolute and relative systematic deviations. The data structure for the F-Al procedure is given in Table 8. The procedure was characterized by LQ value of 0.0038 mg of quercetin and by systematic deviations ranging from - 12.34% to + 0.76%. It is likely that small deviations of blank and gross values are the principal gen- erator of random deviations ranging from ± 0.65% to ± 7.22%. The extensive prevalidation metrological characteristics are summarized in Table 9. 3. 8. Quantitative Analysis of Total Flavonoids in Plantago L. Species The yields of total flavonoids in leaves, stems, and flowers of Plantago species are given in Table 10. The results of F-Al procedure showed that generally the highest content of flavonoids was observed in leaves, compared to stems and flowers. The yields of flavonoids in leaves varied from 0.053% (P coronopus) to 0.131% (P maritima). The maximum flavonoid concentration in stems was 0.065% (P. holosteum subsp. depauperata), while the smallest amount contained sample of P. coronopus (0.008%). Flowers of P. argentea (0.067%) were the most abundant with flavonoids, while the minimum concentration was determined in flowers of P. maritima (0.007%). Generally, the highest content of total flavonoids was determined in above-ground parts (leaves + stems + flowers) of P. argentea (0.221%), while the lowest amount Table 9: Prevalidation characteristics of F-Al procedure for quercetin determination. Parameter F-Al procedure Working range [mg] Information value range [absorbance units] Analyte-signal relationship Calibration function Analytical evaluation function Standard deviation of procedure Limit of detection, LD [mg] Limit of quantitation, Lq [mg]_ 0.240 - 0.024 0.6999 - 0.0594 r = 0.9996 S = 2.85 x x = 0.35 S ± 0.0035 0.0011 0.0038 Groups data Actual [mg] 0.240 0.192 0.144 0.096 0.048 0.024 Found [mg] 0.244 0.194 0.141 0.097 0.040 0.023 Random deviations s^ [mg] ± 0.002 ± 0.001 ± 0.003 ± 0.002 ± 0.002 ± 0.002 srx, [%] ± 0.65 ± 0.77 ± 1.88 ± 2.31 ± 5.19 ± 7.22 Systematic deviations, aAbsolute [mg] + 0.0016 + 0.0016 - 0.0023 - 0.0004 - 0.0059 - 0.0028 Relative [%] + 0.64 + 0.76 - 1.61 - 0.39 - 12.34 - 11.70 Table 10: Content of total flavonoids in different plant organs of Plantago L. species. Plant Total flavonoids (%); X ± SD, n = 3 Leaves Stems Flowers P. altissima 0.095 ± 0.017 0.013 ± 0.001 0.024 ± 0.001 P. argentea 0.110 ± 0.037 0.044 ± 0.002 0.067 ± 0.002 P. coronopus 0.053 ± 0.015 0.008 ± 0.001 0.025 ± 0.002 P. holosteum subsp. depauperata 0.115 ± 0.019 0.065 ± 0.003 0.039 ± 0.002 P. holosteum subsp. holosteum 0.093 ± 0.019 0.047 ± 0.002 0.032 ± 0.002 P. holosteum subsp. scopulorum 0.065 ± 0.018 0.049 ± 0.002 0.011 ± 0.001 P. lagopus 0.094 ± 0.019 0.038 ± 0.001 0.026 ± 0.005 P. maritima 0.131 ± 0.011 0.038 ± 0.001 0.007 ± 0.001 of the examined compounds was established for the above-ground parts of P. coronopus (0.089%). 3. 9. Mathematical/statistical Evaluation of Flavonoid Analysis in Plantago L. Species As regard to content of total flavonoids in leaves, stems and flowers, Principal Component Analysis (PCA) separated investigated Plantago species as it is shown on Figure 1. Table 11: Eigen-vectors of the principal components. Figure 1: PCA of total flavonoids in Plantago L. species. The most similar species were P. holosteum subsp. holosteum and P. lagopus. Higher degree of separation showed samples of P. argentea, P. holosteum subsp. de-pauperata and P. maritima. These three species had the highest contents of total flavonoids. Above-ground parts of P. coronopus contained the smallest amount of examined compounds, so it was also significantly separated from the species in the central part of the PCA scatterplot. The first principal component explains 53.76% of the total variance, the second one 28.14% and the third component explains 18.10% of the variance. Eigen-vec-tors matrix with the loading of each variable in each prin- Variable PC 1 PC 2 PC 3 Flavonoids (leaves) 0.624900 -0.360836 0.692313 Flavonoids (stems) 0.636767 -0.277492 -0.719393 Flavonoids (flowers) 0.451694 0.890391 0.056364 cipal component is presented in Table 11. The highest contribution to the first PC axis gave the content of flavonoids in stems and leaves. The content of flavonoids in flowers contributed the most to the second PC axis. The results of F-Al procedure were also evaluated by using Student's t-test (Table 12) in order to illustrate various distributions of total flavonoids in different plant organs of the same plant species. The greatest difference was established for the specimens of P. maritima, while the statistically insignificant difference was obtained for P. holosteum subsp. scopulorum. In general, statistically significant differences were observed between flavonoid content in stems and flowers of the same species (p < 0.001, except in P. lagopus: p < 0.02). 4. Conclusions Very simple, useful, and informative prevalidation concept for quality control and standardization of analytical procedure was used to obtain prevalidation characteristics of procedure for spectrophotometric determination of flavonoids with AlCl3. Good metrological characteristics obtained for F-Al procedure confirmed the usefulness of the system under study which is characterized by both ideal calibration and analytical evaluation functions, very low limit of quantitation (LQ = 0.0038 mg) and favourably random (from ± 0.65% to ± 7.22%) and systematic (from - 12.34% to + 0.76%) deviations accordant with prevalidation acceptance criteria. Favourable prevalidation characteristics confirm the usefulness of F-Al procedure as a standard method for flavonoids determination in plant material. The results of flavonoids analysis in Plantago species performed by F-Al procedure showed that leaves generally contained the greater amount of flavonoids Table 12: Statistical comparison of total flavonoid content in different plant organs of investigated Plantago L. species using Student's t-test. Plant Probability (p) Leaf-stem Leaf-flower Stem-flower P. altissima < 0.010 < 0.010 < 0.001 P. argentea < 0.050 < 0.200 < 0.001 P. coronopus < 0.010 < 0.050 < 0.001 P. holosteum subsp. depauperata < 0.020 < 0.010 < 0.001 P. holosteum subsp. holosteum < 0.020 < 0.010 < 0.001 P. holosteum subsp. scopulorum < 0.300 < 0.010 < 0.001 P. lagopus < 0.010 < 0.010 < 0.020 P. maritima < 0.001 < 0.001 < 0.001 compared to stems and flowers of investigated plants. The highest concentration was determined in leaves of P. maritima (0.131%), while the highest content in stems and flowers were obtained for the specimens of P. holosteum subsp. depauperata (0.065%) and P. argentea (0.067%), respectively. The highest total flavonoid content in above-ground parts had P. argentea (0.221%), while the lowest amount was determined in the above-ground parts of P. coronopus (0.089%). Multivariate analysis (PCA) of total flavonoids in Plantago species showed that the most similar species were P. holosteum subsp. holosteum and P. lagopus. PCA also pointed out species: P. holosteum subsp. depauperata, P. argentea and P. maritima, with the highest amount of total flavonoids. Student's t-test revealed differences in distributions of total flavonoids in different plant organs of the same plant species. The greatest difference was established for the specimens of P. maritima and the lowest for P. holo-steum subsp. scopulorum. Ultimately, the present study showed that the prevalidation strategy has proven valuable for evaluating the validity of F-Al procedure, which was successfully applied for flavonoids determination in plant material. Moreover, the obtained results of the performed analytical procedure and statistical analysis have contributed to the investigation of the complex genus Plantago. 5. References 1. CPMP/ICH, Working Party on Quality of Medicinal Products, Analytical Validation, 1989, Commission of the European Communities, pp. 1-9. 2. ICH, Harmonized Tripartite Guideline prepared within the International Conference on the Harmonization of Technical Requirements for the Registration of Pharmaceuticals for Human Use (ICH) Q2A, Text on Validation of Analytical Procedures, 1994, Available from http://www.ich.org/ich5q.html. 3. ICH, Harmonized Tripartite Guideline prepared within the International Conference on the Harmonization of Technical Requirements for the Registration of Pharmaceuticals for Human Use (ICH) Q2B, Validation of Analytical Procedures: Methodology, 1996, Available from http://www.ich.org/ich5q.html. 4. Food and Drug Administration, Draft Guidelines on the Validation of Analytical Procedures: Methodology, Federal Register, 1996, 61, 9316-9319. 5. Ph. Hubert, P. Chiap, J. Crommen, B. Boulanger, E. Chapu-zet, N. Mercier, S. Bervoas-Martin, P. Chevalier, M. Laurentie, J. C. Nivet, Anal. Chim. Acta, 1999,391, 135-148. 6. J. W. A. Findlay, W. C. Smith, J. W. Lee, G. D. Nordblom, I. Das, B. S. DeSilva, M. N. Khan, R. R. Bowsher, J. Pharm. Biomed. Anal., 2000, 21, 1249-1273. 7. The United States Pharmacopeia, Section 1225, Validation of compendial methods, United States Pharmacopeial Convention, Rockville, 2002, pp. 2256-2259. 8. M. Thompson, S. L. R. Ellison, R. Wood, PureAppl. Chem., 2002, 74, 835-855. 9. B. Boulanger, P. Chiap, W. Dewe, J. Crommen, Ph. Hubert, J. Pharm. Biomed. Anal., 2003,32, 753-765. 10. V. Grdinic, J. Vukovic, J. Pharm. Biomed. Anal., 2004, 35, 489-512. 11. J. B. Harborne. (ed), The Flavonoids, Advances in Research Since 1986, Chapman and Hall, London, 1994. 12. K. Robards, M. Antolovich, Analyst, 1997,122, 11R. 13. Y. T. Chen, R. L. Zheng, Z. J. Jia, Y. Ju, Free Radic. Biol. Med., 1990, 9, 19-21. 14. D. Pathak, K. Pathak, A. K. Singla, Fitoterapia, 1991, 62, 371-389. 15. A. L. Miller, Alt. Med. Rev., 1996,1, 103-111. 16. T. N. Kaul, E. Jr. Middleton, P. L. Ogra, J. Med. Viral., 1985, 15, 71-79. 17. L. M. Larocca, L. Teofili, S. Sica, M. Piantelli, N. Maggiano, G. Leone, F. O. Raneletti, Blood, 1995,85, 3654-3661. 18. R. L. Singhal, Y. A. Yeh, N. Prajda, E. Olah, G. W. Sledge, G. Weber, Biochem. Biophys. Res. Comm., 1995,208, 425-431. 19. M. A. Pereira, C. J. Grubbs, L. H. Barnes, H. Li, G. R. Olson, I. Eto, M. Juliana, L. M. Whitaker, G. J. Kelloff, V. E. Steele, R. A. Lubet, Cancerogenesis, 1996,17, 1305-1311. 20. S. Caltagirone, F. O. Raneletti, A. Rinelli, N. Maggiano, A. Colasante, P. Musiani, F. B. Aiello, M. Piantelli, Am. J. Resp. Cell. Mol. Biol., 1997,17, 51-59. 21. S. O. Keli, M. G. Hertog, E. J. Feskens, D. Kromhout, Arch. Intern. Med., 1996,156, 637-642. 22. R. Taskova, Lj. Evstatieva, N. Handjieva, S. Popov, Z. Naturforsch., 2002, 57c, 42-50. 23. A. B. Samuelsen, J. Ethnopharm., 2000, 71, 1-21. 24. S. A. Kawashty, E. Gamaleldin, M. F. Abdalla, N. A. M. Saleh, Biochem. Syst. Ecol., 1994, 22, 729-783. 25. S. R. Jensen, C. E. Olsen, K. Rahn, J. H. Rasmussen, Phytochemistry, 1996, 42, 1633-1636. 26. R. Jurisic, Z. Debeljak, S. Vladimir-Knezevic, J. Vukovic, Z. Naturforsch., 2004, 59c, 27-31. 27. S. Kulevanova, M. Stefova, T. Stafilov, Acta Pharm., 2000, 50, 29-37. 28. Z. Males, I. Zuntar, B. Nigovic, M. Plazibat, V. Bilusic Vundac, Acta Pharm. 2003, 53, 139-144. 29. M. Tomczyk, J. Gudej, Z. Naturforsch., 2003, 58c, 762-764. 30. J. Gudej, M. Tomczyk, Arch. Pharm. Res., 2004, 27, 11141119. 31. B. Christ, K. H. Müller, Arch. Pharm., 1960, 293, 1003. 32. Deutsches Arzneibuch, 10. Ausgabe, Deutscher Aptheker Verlag, Stuttgart, Govi Verlag, GmbH, Frankfurt, 1991. 33. R. D. Petry, G. G. Ortega, W. B. Silva, Pharmazie, 2001, 56, 465-470. 34. S. Sharma, Applied Multivariate Techniques, John Wiley & Sons, Inc., 1996. 35. B. G. M. Vandeginste, D. L. Massart, L. M. C. Buydens, S. De Jong, P. J. Lewi, J. Smeyers-Verbeke, Handbook of Chemometrics and Qualimetrics: Part B., Elsevier Science, Amsterdam, 1998, pp. 349-380. 36. K. McGarigal, S. Cushman, S. Stafford, Multivariate Statis- tics for Wildlife and Ecology Research, Springer-Verlag, 2000, New York. 37. M. Höft, S. K. Barik, A. M. Lykke, Quantitative Ethnobo-tany: Applications of Multivariate and Statistical Analyses in Ethnobotany, People and Plants working paper 6, UNESCO, Paris, 1999, p. 46. 38. J. H. Mook, J. Haeck, J. van der Toorn, P. H. van Tienderen, in Plantago: A Multidisciplinary Study, eds. P. J. C. Kuiper, M. Bos, Springer-Verlag, 1992, Berlin, 69-87. 39. J. N. Miller, J. C. Miller, Statistics and Chemometrics for Analytical Chemistry, Pearson Education Limited, 2000, Essex. 40. IUPAC, Compendium of analytical nomenclature, Definitive rules, 1997, Available from http://www.iupac.org/analytical _compendium. 41. D. L. Massart, B. G. M. Vanderginste, L. M. C. Buydens, S. De Jong, P. J. Lewi, J. Smeyers-Verbeke, Handbook of Chemometrics and Qualimetrics: Part A, Elsevier, Amsterdam, 1997. 42. G. Gottschalk, Z. Anal. Chem., 1975, 276, 81-95. Povzetek V delu je opisan postopek validacije in uporaba spektrofotometrične metode UV/Vis z aluminijevim trikloridom kot reagentom za določevaje flavonoidov v rastlinskih vzorcih. Za kontrolo kakovosti in standardizacijo analiznega postopka ter potrditev zanesljivosti metode je bila izvedena hitra in enostavna validacija na osnovi matematično-statističnega testiranja. Ugodni parametri validacije so potrdili primernost metode za določevanje flavonoidov z AlCl3 (F-Al metoda), ki je bila tudi uspešno uporabljena za določevanje flavonoidov v listih, steblih in cvetovih trpotcev Plantago L. Rezultati so pokazali različno vsebnost flavonoidov v različnih deli rastlin vzorčenih na Hrvaškem (listi: do 0,13 %; stebla: do 0,07 % in cvetovi: do 0,07 %). Rezultati določevanja flavonoidov so bili za proučevanje različnih taksonov iz rodu Plantago L. statistično ovrednoteni z metodo analize glavnih osi in Studentovega t-testa.