49 Članek prispel / Received 4. 2. 2020 Članek sprejet / Accepted 9. 4. 2021 Abstract Purpose: The objective of this study was to introduce the GeneXpert Ultra cartridge-based BCR-ABL diagnostic test into routine molecular monitoring of mi- nimal residual disease for chronic myelo- id leukemia (CML) patients who are on tyrosine kinase inhibitors (TKIs). Methods: BCR-ABL1 transcript quantification was performed on 100 peripheral blood samples from 33 CML patients at different stages of the disease using automated GeneXpert BCR-ABL Ultra assay and following a standardized laboratory protocol (SP). Results: Analysis results were expressed as % on the International Scale (IS) for Izvleček Namen: Cilj študije je bil uvesti dia- gnostični test BCR-ABL, ki temelji na GeneXpert Ultra sistemu, za rutinsko molekularno spremljanje minimalne preostale bolezni (MRD) pri bolnikih s kronično mieloidno levkemijo (CML), ki so na zaviralcih tirozin kinaze (TKI). Metode: Kvantifikacija transkriptov BCR-ABL1 je bila izvedena v 100 vzor- cih periferne krvi 33 bolnikov s CML v različnih fazah bolezni z avtomati- ziranim GeneXpert BCR-ABL Ultra testom (Cepheid, ZDA) in s standardizi- ranim laboratorijskim protokolom (SP). Rezultati: Rezultati obeh analiz so bili izraženi kot % na mednarodni le- Ključne besede: kronična mieloična levkemija, fuzijski gen BCR-ABL1, GeneXpert, minimalna preostala bolezen, molekularni odziv. Key words: chronic myeloid leukemia, BCR- ABL1 fusion gene, GeneXpert, minimal residual disease, molecular response Natančnost in hitrost poročanja o molekularnem odzivu pri KML bolnikih z in vitro diagnostičnim testom Xpert BCR-ABL Accuracy and speed of molecular response reporting with Xpert BCR-ABL Ultra in vitro diagnostic test in CML patients Avtor / Author Špela Stangler Herodež1,2, Nadja Kokalj-Vokač1,2, Zlatko Roškar3, Mojca Dreisinger3 Ustanova / Institute 1Univerzitetni klinični center Maribor, Klinika za ginekologijo in perinatologijo, Laboratorij za medicinsko genetiko, Maribor, Slovenija; 2Univerza v Mariboru, Medicinska fakulteta, Katedra za molekularno biologijo, Maribor, Slovenija; 3Univerzitetni klinični center Maribor, Klinika za interno medicino, Oddelek za hematologijo in hematološko onkologijo; 1University Medical Centre Maribor, Clinic for Gynecology and Perinatology, Laboratory of Medical Genetics, Maribor, Slovenia; 2University of Maribor, Faculty of Medicine, Department for Molecular Biology, Maribor, Slovenia; 3University Medical Centre Maribor, Clinic for Internal Medicine, Department for Haematology and Haematological Oncology, Maribor, Slovenia. Laboratorijska študija / Laboratory study ACTA MEDICO-BIOTECHNICA 2021; 14 (1): 25–37 50 Laboratorijska študija / Laboratory study ACTA MEDICO-BIOTECHNICA 2021; 14 (1): 25–37 INTRODUCTION Chronic myeloid leukemia (CML) is a clonal malignant disease of pluripotent stem cells directed towards granulopoiesis. According to the World Health Organization (WHO), it is classified as chronic myeloproliferative disease with an annual incidence of 1.6/100,000, where one out of every 62,500 men and women will be diagnosed with CML during their lifetime (1). The average age at diagnosis is 50–60 years, although the disease can occur during all stages of life, even in childhood. Exposure to ionizing radiation is the only known risk factor for the occurrence of the disease. CML accounts for 20–25% of all leukemia cases. It is Naslov za dopisovanje / Correspondence Doc. dr. Špela Stangler Herodež, univ. dipl. inž. kem. teh., Univerzitetni klinični center Maribor, Klinika za ginekologijo in perinatologijo, Laboratorij za medicinsko genetiko, Ljubljanska 5, 2000 Maribor, Slovenija Telefon +386 23212737 Fax +386 23212755 E-pošta: Spela.STANGLERHERODEZ@ukc-mb.si stvici (IS) in so bili ocenjeni z moleku- larnim odzivom (MR) v skladu z labora- torijskimi priporočili za oceno globokih molekulskih odzivov po zdravljenju kronične mieloične levkemije. Njihova primerjava je pokazala, da ni bilo stati- stično pomembne razlike v rezultatih, iz- raženih kot % na IS, p = 0,098. Vzorce glede na ocenjeno MR smo razdelili v 6 skupin (MR1, MR2, MR3, MR4, MR4.5 in MR5 ). 74 od 100 vzorcev je imelo z obema metodama popolnoma enak MR. 12 vzorcev je imelo višji MR, pri- dobljen z GeneXpertom, medtem ko je imelo 14 vzorcev nižji MR v primerjavi s SP. Pri 3 od 74 vzorcev se je GeneXpert razlikoval od SP za dva MR razreda, medtem ko se je pri ostalih razlikoval za en MR razred. Analiza glavnega mo- lekularnega odziva (MMR) je pokazala, da smo dosegli 82 % ujemanje. V osmih vzorcih so bile z GeneXpertom zaznane minimalne količine kopij BCR-ABL1, medtem ko jih SP ni zaznal. Ključno je, da pri nobenem izmed bolnikov re- zultat, pridobljen z GeneXpertom, ni vplival na drugačno klinično odločitev o zdravljenju s TKI kot rezultat, ki je bil pridobljen po SP. Zaklju~ek: Glede na ugotovitve lah- ko Xpert BCR-ABL Ultra test šteje- mo kot koristno in zelo hitro klinično orodje za molekularno spremljanje bolnikov na TKI. BCR-ABL1 real-time PCR quantifica- tion and estimated using molecular re- sponse (MR) according to the laboratory recommendations for scoring deep MRs following CML treatment. A compari- son of the two methods showed that there was no statistically significant difference in the results expressed as % on the IS (p = 0.098). Samples were divided into six groups (MR1, MR2, MR3, MR4, MR4.5, and MR5) according to the esti- mated MR. A total of 74/100 samples had a completely identical MR with both methods, 12/100 samples had a higher MR obtained with GeneXpert, and 14/100 samples had a lower MR compared to the SP. In addition, 2/74 samples had a different MR for two clas- ses obtained with GeneXpert compared to the SP, while the others 72/74 sam- ples differed in one MR class. Analysis of the major molecular response resulted in 82% agreement. Minimal amounts of BCR-ABL1 copies were detected with GeneExpert in eight samples while the SP did not detect them. None of the GeneXpert results had any effect on the clinical decision for treatment with TKIs obtained using the SP. Conclusion: The Xpert BCR-ABL Ul- tra assay can be considered a useful and very fast clinical tool for molecular follow- -up of patients on TKIs. 51 Laboratorijska študija / Laboratory study ACTA MEDICO-BIOTECHNICA 2021; 14 (1): 25–37 the first malignant disease in which a change leading to its development was recognized at a molecular level (2). It is caused by the formation of the Philadelphia chromosome, a reciprocal translocation of the chromosomal parts between chromosomes 9 (9q34.1) and 22 (22q11.2), resulting in the formation of the BCR-ABL1 fusion gene. The resulting fusion oncogene carries the BCR-ABL1 (tyrosine kinase) protein that is responsible for the development of CML because it maintains the activity of the signaling pathway in the cell, thereby accelerating the proliferation of stem cells, interfering with maturation, and preventing apoptosis. Genetically modified cells overwhelm the bone marrow. Genetic instability of the cell clone may cause resistance to the treatment and progression of the disease to secondary acute leukemia. Therefore, BCR- ABL1 can be employed as a specific leukemic marker for minimal residual disease (MRD) follow-up (3). After treatment with tyrosine kinase inhibitors (TKIs), BCR- ABL1 quantification may predict the outcome in CML patients, especially in those who achieve a complete cytogenetic response (4). Monitoring of BCR-ABL transcript levels with real- time quantitative polymerase chain reaction (qRT- PCR) in patients on TKI therapy is the standard of care in the management of CML. Furthermore, it is very important to achieve sensitivity of 4.5 logs below baseline in patients who are considered for TKI discontinuation (5). In order to assess the treatment response, it is essential to ensure that variability in testing methodologies is tightly controlled. This was the cause of BCR-ABL molecular measurement standardization across laboratories, which introduced the International Scale (IS) defined by WHO (5). The GeneXpert system utilizes microfluidics to directly process peripheral blood samples in a cartridge for RNA extraction, reverse transcriptase PCR (RT-PCR), and fluorescence detection in an “all-in-one” self-contained instrument (6). The GeneXpert BCR-ABL system uses ABL1 as its control gene, which is recommended for molecular response (MR) follow-up. It calculates the BCR-ABL1/ABL1 ratio using the delta-crossing threshold (Ct) comparative-quantitative method (7). Jobbagy et al. and Winn-Deen et al. performed an early comparison of the GeneXpert assay with standard qRT- PCR assay and found that the system had the potential to translate into clinical practice despite the differences in quantitative methods (8, 9). However, the system did not report BCR-ABL1 values on the IS and result interpretation and inter-laboratory comparison were hindered. Enjeti et al. developed a conversion factor specific to GeneXpert assay and used it to adjust the BCR-ABL1/ABL1 ratio to the IS. A batch-specific standard curve and an efficiency value were also introduced to ensure cartridge lot-specific calibration (10). This very important quality control assay improvement enables inter-laboratory comparison. It also achieved an accurate measurement of 0.01% of BCR-ABL1 on the IS, a level sufficient to detect MR4 (molecular response at 4-log reduction) (11, 12). In our study, we introduced the GeneXpert Ultra cartridge-based BCR-ABL diagnostic test into routine molecular monitoring of MRD for CML patients who are on TKIs. MATERIALS AND METHODS Study sample Peripheral blood samples were obtained from CML patients on TKI therapy referred to our laboratory of Medical Genetics from the Department of Haematology and Haematological Oncology, Clinic for Internal Medicine, University Medical Centre Maribor, Maribor, Slovenia for monitoring of BCR-ABL transcript levels. All samples were from individuals of Caucasian descent who were residents of different geographical areas of east Slovenia. The study was approved by the ethical committee (No. UKC-MB-KME-24-08/17). Informed consent was obtained from all participating individuals. Assay design An automated GeneXpert BCR–ABL Ultra car- tridge System (Cepheid Europe S.A.S., Vira-Solelh, Maurens-Scopont, France) consists of an instrument and a personal computer with software for running tests and viewing results. It simplifies generation of very accurate and reproducible results using integra- tion of nucleic acid extraction, target amplification, and quantification of the BCR-ABL e13a2/b2a2 and e14a2/b3a2 transcripts relative to ABL in peripher- al blood samples. Disposable GeneXpert cartridges 52 Laboratorijska študija / Laboratory study ACTA MEDICO-BIOTECHNICA 2021; 14 (1): 25–37 contain the RT-PCR and nested PCR reagents and host the RT-PCR and nested PCR processes. Cross- contamination concerns are minimized because the cartridges are self-contained. Assay procedure Peripheral venous blood samples were collected in standard vacutainer tubes containing ethylenedi- amine tetraacetic acid (EDTA) anticoagulant. A total of 4 mL of whole peripheral blood was mixed with 100 µL of proteinase K (Cepheid Europe S.A.S.). Af- ter 1 min of incubation at room temperature, 2.5 mL of lysis reagent (Cepheid Europe S.A.S.) was added to inactivate nucleases in order to release nucleic acids from the cells, followed by vortexing and two incuba- tion periods of 5 min at room temperature. After the second incubation, 1 mL of lysate was transferred to a new tube and mixed with 1.5 mL of lysis reagent. Af- ter 10 min of incubation at room temperature, 2 mL of pure ethanol (Merck KgaA, Darmstadt, Germany) was added to the sample and 4.5 mL of mixture was transferred to the test cartridge. Washing reagent (Ce- pheid Europe S.A.S.) was then added to the designat- ed port in the test cartridge, the lid was closed, and the cartridge was loaded into the GeneXpert System. The GeneXpert System isolates the total RNA and runs a nested RT-PCR. Total assay time was approximately 2 h and 20 min. Assay quality control Each GeneXpert BCR-ABL Ultra kit lot is standard- ized using in-house secondary standards calibrated ac- cording to the WHO (Figure 1). 13 Figure 1. Sample of the BCR-ABL ΔCt standard curve. EΔCt=1.96 (lot-specific efficiency value) is derived from the slope of the standard curve; IS-SF = 1.26 (lot-specific scaling fac- tor) corrects the quantitative output of the assay to the international scale established based on WHO Interna- tional Genetic Reference Panel. Figure 1: Sample of the BCR-ABL ΔCt standard curve. EΔCt=1.96 (lot-specific efficiency value) is derived from the slope of the standard curve; IS-SF = 1.26 (lot-specific scaling factor) corrects the quantitative output of the assay to the international scale established based on WHO International Genetic Reference Panel. 14 Figure 2. Calibration to WHO IS. A lot-specific conversation factor is applied to each lot (Figure 2). Statistical Analysis MRs were compared using the t-test, where p ≤ 0.05 was considered statistically significant. RESULTS BCR-ABL1 transcript quantification was performed on 100 peripheral blood samples from 33 CML patients at different stages of the disease in two different laboratories. The first method used the automated GeneXpert BCR- ABL Ultra assay (Cepheid Europe S.A.S.), while the second used a standardized laboratory protocol (SP). The blood samples were collected over three years in the period from 2017 to 2020. The negative control group included 14 blood samples obtained from patients with hematological disorders 53 Laboratorijska študija / Laboratory study ACTA MEDICO-BIOTECHNICA 2021; 14 (1): 25–37 DISCUSSION A major advantage of the GeneXpert platform is its simplified work flow, which takes approximately 2 h and 20 min. The system thus provides a faster turnaround, reduces hands-on time, and requires less technical expertise compared to the standard BCR- ABL1 qRT-PCR assay. The GeneXpert automates both the pre-analytical and analytical phases. The risk of contamination with the equipment or with other samples is insignificant because the assay is performed inside a closed cartridge. All of these factors support the choice of GeneXpert for BCR-ABL1 quantification and CML monitoring. Cepheid has also introduced the GeneXpert BCR-ABL Ultra, a high-sensitivity assay, which demonstrates a limit of detection reaching MR4.5 and below (13). BCR-ABL assays that are sensitive in their measurement of deep-level response may aid in the identification of potential candidates for treatment discontinuation. GeneXpert BCR-ABL Ultra detects the most common BCR-ABL transcripts below MR4.5 or 0.0032%, which is widely accepted as the clinical threshold that defines candidates who can safely discontinue TKI therapy (13). This important GeneXpert BCR-ABL1 assay improvement has led our laboratory to introduce the system into routine molecular monitoring of MRD for CML patients who are on TKIs, because the ability to consistently detect low levels of BCR-ABL transcripts in CML patients is important for the assessment of treatment outcomes in patients on TKI therapy. For that reason, GeneXpert BCR-ABL Ultra results in the present study were compared to the SP results. The concordance analysis showed that 74% of samples agreed at the same logarithmic interval and had a similar distribution of GeneXpert disagreement in values above and below those obtained using the SP. Analysis by intervals did not show any significant difference in concordance (p = 0.098). Analysis of MMR showed an 82% agreement. MR was demonstrated to be properly determined using GeneXpert when compared to the SP. Our results also highlight the fact that GeneXpert was able to detect a low number of copies in cases that turned out to be negative when determined using the SP. This indicated a higher sensitivity as a result of the fact that GeneXpert provides a more sensitive Table 1. Concordance between standardized laboratory protocol (SP) and GeneXpert Ultra assay results at different logarithmic intervals and overall. MR (%IS) Number of samples at the interval Concordance between methods All samplesSP GeneXpert Ultra MR1 (>1) 7 5 5/7 (71 %) MR2 (≤1) 5 (50.0) 34 (18.7) 6/6 (100 %) MR3 (≤0.1) 0 (0.0) 22 (12.1) 13/16 (81 %) MR4 (≤0.01) 0 (0.0) 13 (7.1) 5/7 (71 %) MR4.5 (≤0.0032) 10 (100.0) 182 (100.0) 4/6 (67 %) MR5 (≤0.001) 0 (0.0) 13 (7.1) 41/50(82 %) Not detected 10 (100.0) 182 (100.0) 0/8 (0 %) Total 0 (0.0) 13 (7.1) 74/100 (74 %) unrelated to BCR-ABL gene fusion. The negative predictive value was 100%. A total of 74 of 100 (74%) samples had a completely identical MR using both methods (Table 1). Furthermore, 12 of 100 samples had a higher MR obtained with GeneXpert, while 14 of 100 samples had a lower MR compared to the SP. Analysis by intervals did not show any significant difference in concordance (p = 0.098). Furthermore, only two of 74 (2.7%) samples had a different MR for two classes obtained with GeneXpert compared to the SP. The others 72 samples differed in one MR class. Analysis of the major molecular response (MMR) resulted in 82% agreement (Table 1). Minimal amounts of BCR-ABL1 copies were detected in eight samples that were not detected by the SP (Table 1). None of the GeneXpert results had any effect on the clinical decision for treatment with TKIs obtained using the SP. 54 Laboratorijska študija / Laboratory study ACTA MEDICO-BIOTECHNICA 2021; 14 (1): 25–37 assay (nested PCR) that is able to detect reduction of up to 5 log in BCR-ABL1 transcript concentrations. The negative control group also had a high specificity (100%) using GeneXpert. It can therefore be concluded that the GeneXpert BCR-ABL Ultra assay is a reproducible and robust alternative to laboratory-developed assays. Its ease of use may allow some laboratories to offer BCR-ABL testing to the patients they serve. It is also important that the short assay time enables same-day results for treating clinicians. Such fast sample turnaround can reduce patient anxiety and allow a faster treatment adjustment if there is an increase in BCR-ABL concentration. ACKNOWLEDGEMENTS This work was supported by the IRP-2017/01-05 internal research project of University Medical Centre Maribor, Maribor, Slovenia. The authors are grateful to all patients and individuals in the study who made this work possible and would especially like to thank the members of the Department for Haematology and Haematological Oncology, Clinic for Internal Medicine, University Medical Centre Maribor, Maribor, Slovenia and the members of the Clinical Department for Hematology, Clinic for Internal Medicine, University Medical Centre Ljubljana, Ljubljana, Slovenia for their helpful contributions to this study. REFERENCES 1. Siegel RL, Miller KD, Jemal A. Cancer statistic, 2019. CA Cancer J Clin 2019; 69: 7-34. 2. Höglund M, Sandin F, Simonsson B. Epidemiol- ogy of chronic myeloid leukaemia: an update. Ann Hematol 2015; 94(Suppl 2): S241-7. 3. Branford S, Haughes TP. Practical considerations for monitoring patients with chronic myeloid leu- kemia. Semin Hematol 2010; 47: 327-34. 4. Hughes TP, Hochhaus A, Branford S, Müller MC, Kaeda JS, Foroni L. Long-term prognostic signifi- cance of early molecular response to imatinib in newly diagnosed chronic myeloid leukemia: an analysis from the International Randomized Study of Interferon and ST1571 (IRIS). Blood 2010; 116: 3758-65. 5. Day GJ, Lockwood C, Payton JE, Uy GL, Schiffer C, Feldman G et al. Development of Xpert® BCR-ABL Ultra, an Automated and Standard- ized Multiplex Assay with Required Performance Characteristics for BCR-ABL1 Quantitative Mea- surement on an International Reporting Scale. Blood 2015; 126: 2793. 6. Dufresne S, Belloni D, Levy N. Tsongalis G. Quan- titative assessment of the BCR-ABL transcript us- ing the Cepheid Xpert BCR-ABL monitor assay. Arch Pathol Lab Med 2007; 131: 947-50. 7. Beillard E, Pallisgaard N, Van der Velden VH, Bi W, Dee R. Evaluation of candidate control genes for diagnosis and residual disease detection in leukemic patients using real-time quantitative reverse-transcriptase polymerase chain reaction (RQ-PCR) – a Europe Against Cancer Program. Leukemia 2007; 17: 2474-86. 8. Jobbagy Z, Van Atta R, Murphy K, Eshleman J, Gocke C. Evalution of the cepheid Gene Xpert BCR-ABL assay. J Mol Diag 2007; 9: 220-7. 9. Winn-Deen E, Helton B, Van Atta R, Wong W, Peralta J, Wang J et al. Development of an inte- grated assay for detection of BCR-ABL RNA. Clin Chem 2007; 53: 1593-600. 10. Enjeti A, Granter N, Ashraf A, Fletcher L, Bran- ford S, Rowlings P et al. A longitudinal evaluation of performance of automated BCR-ABL1 quan- titation using cartridge-based detection system. Pathology 2015; 47: 570-4. 11. Cayuela J, Macintyre E, Darlington M, Abdelali R, Fund X, Villarese P et al. Cartridge-based auto- mated BCR-ABL1 mRNA quantification: solving the issues of standardization, at what cost? Hae- matological 2011; 96: 664-71. 12. O'Dwyer M, Swords R, Nagler A, McMullin A, Le Coutre P, Langabeer S et al. Nilotinib 300 mg 55 Laboratorijska študija / Laboratory study ACTA MEDICO-BIOTECHNICA 2021; 14 (1): 25–37 bid as front-line treatment of CML: prospective analysis of the Xpert BCR-ABL monitor system and significance of 3-month molecular response. Leuk Res 2014; 38: 310-5. 13. Ferrand C, Deschamps M, Levitas A, Lockwood C, Payton JE, Uy GL et al. Xpert® BCR-ABL Ultra, a High Sensitivity Assay, Demonstrates a Limit of Detection Reaching MR4.5 and below. Blood 2017; 130: 1610.