Advances in Production Engineering & Management Volume 12 | Number 1 | March 2017 | pp 88-96 https://doi.Org/10.14743/apem2017.1.242 ISSN 1854-625G Journal home: apem-journal.org Original scientific paper Automatic high resolution measurement set-up for calibrating precise line scales Klobucar, R.a*, Acko, B.a aFaculty of mechanical engineering, University of Maribor, Maribor, Slovenia A B S T R A C T A R T I C L E I N F O This paper presents a high resolution measurement set-up developed for calibrating precise line scales with measurement uncertainty of less than 0.1 |m over a total length of 500 mm. The system integrates a numerically controlled multi-axis stage, a laser interferometer, and a vision system for detecting line position. The measurement and the analysis processes are completely automated in order to minimize manual labour during the calibration process, but also increase the calibration accuracy. Increasing calibration accuracy leads up to better quality of industrial measurements which is required by modern precision industry. When designing this set-up, special attention was paid to the alignment of the measurement object in the measurement direction, considering the focus of the camera. The aim of this alignment was to reduce Abbe errors in 2 axes to negligible level. In addition, all uncertainty contributions have been determined and evaluated by performing extended experiments in specific measurement conditions. These contributions are presented in the uncertainty budget. The metrological capabilities of the presented measurement set-up were verified by some practical test measurements. Selected results of these measurements are presented in the article. This set-up will primarily improve a standard base for calibration of optical measuring devices. The use of the optical standards in the industry is constantly growing. Indirect users of the results of this research will be all manufacturers of precise products such as automotive and other industries. Keywords: Measurement Line scales High resolution measurements Measurement uncertainty *Corresponding author: rok.klobucar@um.si (Klobucar, R.) Article history: Received 15 September 2016 Revised 27 February 2017 Accepted 1 March 2017 © 2017 PEI, University of Maribor. All rights reserved. 1. Introduction Line scales and grid plates are most common measurement standards for assuring traceability of optical measuring equipment, such as microscopes, profile projectors and digital "vision" systems. They are widely used in industry and research laboratories as calibration standards [8]. Line scales are important material standards of length, used for accurate positioning or measurement in one, two or three dimensions [1]. Industry is demanding more and more rapid and accurate dimensional measurements on diverse mechanical parts. In recent years, optical coordinate measuring machines (CMMs) having made substantial progress and are now often used for such applications. Especially CMMs equipped with imaging capabilities are frequently used for fast, non-contact measurements. Strong competition among manufacturers of such instruments and the demand for sub-micron accuracy has led to standardised tests, which are aimed for comparing and validating instrument performance [2]. Calibration, verification and also error correction of 88 Automatic high resolution measurement set-up for calibrating precise line scales optical CMMs are mainly based on measurements using reference line scales or two-dimensional grid plates [3]. There are very many commercially available line scales made of different materials (steel, brass, invar, glass, quartz, zerodur). They can vary in length from below 1 mm to more than 1 m, and have resolutions (pitch) from below 10 nm to 1 cm. They can be calibrated by using different measurement set-ups, depending on their length and precision. Set-ups for calibrating high-precision line scales normally involve a microscope with an optical sensor for capturing and analysing the image of a line marker and a laser interferometer as a traceable measurement standard [4-6]. This paper introduces a measuring system developed for calibrating line scales with measurement uncertainty of less than 10 nm over a total length of 500 mm. The system integrates a numerically controlled multi-axis stage, a laser interferometer, and a vision system for detecting line position. With this measurement set-up, Abbe errors can be reduced to negligible levels. Abbe error occurs when the measurement point of interest is spatially separated laterally from the actual measuring scale location (reference line or axis of measurement), and when angular error motions exist in the positioning system. Abbe error causes measured displacement to appear longer or shorter than the true displacement, depending on the direction of angular motion. Spatial separation between the measured point and reference line is known as an Abbe offset [7]. Since it is possible to put optical components very close to each other, the air dead path error of the laser interferometer is also negligible. Software for detecting the middle of the lines is based on earlier design from 2009 [9]. In order to improve performance and to achieve better uncertainty, the software was improved and automatized, mechanics and optics have been redesigned and several uncertainty components better characterized. This facility will replace the old measurement facility that is manually operated. A model for evaluating uncertainty and the uncertainty budget for the demonstrated measuring system is presented in the paper. 2. Measurement system 2.1 Numerically controlled multi-axis stage Numerically controlled multi-axis stage was designed and manufactured by Newport-micro controle [10] for the Laboratory for Production Measurement (LTM), University of Maribor, Slovenia. It is shown in Fig. 1, while its schematic diagram is shown in Fig. 2. The measuring system designed by the LTM is intended to perform 2D measurements on various objects, which will be fixed on a measuring table. This research is focused in 1D measurements. Newport's solution is based on a HybrYX-G5 stage featuring a ceramic carriage which freely slides in X and Y axes on a precision lapped granite reference plate, using proprietary pressure-vacuum air bearing design. The carriage is guided along Y axis by a rigid ceramic beam. The beam is supported and guided at each end with the ball bearing carriages of X axis. Both X and Y axes are motorized with linear actuators and include linear glass encoders. Both linear scale glass encoders are LIDA403 made by Heidenhain, length 440 mm for X axis and length 1250 mm for Y axis. This positioning system is equipped with the optional Z-Tip-Tilt and Theta stage. It is motorized by precision actuators equipped with miniature DC servo motors. The position system is built on a heavy granite table, which features a precision reference plane for the air bearing carriage. In addition, the granite structure includes a granite gantry allowing the fixation of a vertical motorized translation stage, which accommodates the measuring sensor, depending on an application. Two rails mounted on the gantry equipped with sliding carriages provide manual rough positioning capability. The granite structure is set on a heavy welded frame equipped with ND40 passive isolators. The position system has one XPS motion controller including dedicated drivers for the five DC motors and the three linear motors. The X axis is dual axis composed of X1 and X2 actuators mounted on the flanks at both ends of the granite table. X1 and X2 carriages move the ceramic beam of the Y axes. The X axis travel Advances in Production Engineering & Management 12(1) 2017 89 Klobučar, Acko range is +/- 175 mm. It has a linear encoder with 5 nm resolution. The maximal speed is 300 mm/s., while the repeatability is 500 nm. The Y axis is a horizontal translation stage using air bearing technology and linear motor. The guiding beam of Y axes sub-assembly is attached to the carriages of the X1 and X2 short travel translation actuators. The guideways involve two parts: a large ceramic carriage and a Y-axis guide ceramic L-shape beam. The ceramic carriage freely slides in X and Y axes using pressure-vacuum air bearing design. The reference planes for the air bearing carriage are the precision lapped surface of the granite table and the vertical slide of the Y rigid ceramic beam. The Y axis travel range is +/- 500 mm. It contains a linear encoder with 5 nm resolution. Maximum speed is 600 mm/s., while the repeatability is 100 nm. The ZTT Theta stage is mounted on the air bearing carriage; it is equipped at the top with the measuring platform, which allows installing measured objects. All axes are driven by the high precision motorized linear actuators. Three actuators are set in upright position with a travel range of 10 mm. One actuator is set horizontally to achieve the theta Z movement with a travel range of +/- 1°. All four actuators are equipped with a miniature DC motor made by Fulhaber. Z axis is a vertical motorized translation stage on the middle carrier. It is a Newport catalogue stage reference M-IMS100V. The Z axis travel range is 100 mm, the minimum incremental motion is 0.3 ^.m, the encoder resolution is 0.1 ^m, the maximum speed is 20 mm/s, while the repeatability is ± 0.5 ^.m. Fig. 1 Measurement set-up for calibrating line scales Measuring beam LASER HEAD Fig. 2 Schematic diagram of measurement set-up for calibrating line scales 90 Advances in Production Engineering & Management 12(1) 2017 Automatic high resolution measurement set-up for calibrating precise line scales 2.2 Laser interferometer Laser interferometer position measurement systems provide very precise position or distance information. Our system consists of a laser head HP 5528A, Agilent module 55292A, and a variety of optical components and accessories such as material sensors and air sensor. The basic system measures linear displacement The system uses the wavelength of light from a low-power helium-neon laser as a length standard. We normally set the resolution of the laser interferometer to 10 nm. Special mounting elements were constructed for the optics used for calibration. A schematic diagram of the laser interferometer and the position of the optics are shown in Fig. 2. The position of the optical elements and the moving parts are set in such way, that the measured object axis is set in the line with the centre of the laser linear retroreflector. It allows the Abbe errors to be reduced to negligible levels even for the most demanding dimensional metrology tasks [11]. The linear interferometer is fixed on the fixed part of the stage, while the retroreflector is placed on the moving table in the horizontal direction. Only the measured object is placed on the moving table that can move in the vertical direction. 2.3 Vision system for detecting line position The vision system for detecting line position consists of a zoom microscope and a CMOS digital camera. The camera is connected to the computer via USB 3.0 port. The CMOS camera gets the images of the line scale and sends them to the computer software. The software analyses the images and determinates the middle of the line in the measured window as shown in Fig. 3, which is defended by the operator. The software calculates the distance in pixels from the reference position, which is marked with the blue line, to the middle of the measured line that, is marked with the red line. The image positioning screen with the measurement window, reference and measured line is presented in Fig. 3. The CMOS digital camera takes 15 monochrome images per second in resolution 2592 x 1944 pixels. The software analyses the images in real time. The distance calculated in pixels is transformed into micrometers [12]. The software for calculation the distance between lines is more detailed presented by the authors in paper [9]. Fig. 3 Screen image of the vision system for detecting line position 2.4 Measurement procedure Line scales are material measures made of glass, steel or other material, on which dimensions are marked with line marks. Since the materials have quite different temperature expansion coefficients, which are in many cases notexactly known, they are stabilized before calibration in the climatic room at 20 °C for 24 hours. The measurement system must be adjusted and initialized before the measurement The line scale should be positioned under the camera by moving the measurement table. The measuring system, laser interferometer, vision system for detecting line position and line scale should be adjusted according to the measuring direction, which is defended by the movement of the horizontal direction of the table. The line scale should be fixed on the adjustment table under the camera. The camera is fixed. It is adjusted in such a way, that the focus of the camera intersects the centre of the laser linear retroreflector. The camera is focused on the lines of the line scale by moving the adjustment table (ZTT Theta stage) Advances in Production Engineering & Management 12(1) 2017 91 Klobučar, Acko in the vertical direction. With this procedure, the line scale axis, the images of the lines and the centre of the laser linear retroreflector with the laser beam are positioned in the measuring direction. The image processing software should be initialized [9]. With the improved software, connection to the numerically controlled multi-axis stage and connection to the laser head, automatic measurements of distances are possible. The measurement goal is to measure distances between the reference line and the chosen measuring lines on the line scale automatically. Measured data are saved into the file where they are ready for further processing. 2.5 Principle for minimising Abbe error In respect to the movement parts of the measuring system, the system can be divided into three different parts. The first part is the fixed part ant it includes the laser head, the linear interferometer and the USB 3.0 camera. The second part is the moving part in the horizontal direction and it includes the linear retroreflector, the moving table and the measured object. The third part is the moving part in the vertical direction and it includes the moving table and the measured object The third part is used only for adjusting the line scale in respect to the measuring object direction. The camera is in the fixed distance, so that the focus of the camera is on the top of the moving table. This means that the theoretical measured lines that could be on the top of the table, are in the focus of the camera and simultaneously with the laser optics. Only the moving table with the line scale could be moved in the vertical direction. This procedure allows us moving the lines of the line scale into the focus of the camera and in the same time into the centre of the laser optics. With this procedure, the minimum Abbe error can be achieved. One of the most important sources of error in dimensional measurement is Abbe error. Abbe error consists of an Abbe offset and a small angular variation due to pitch and yaw of the stage that moves the measuring table with the measured object and the linear retroreflector. This component is caused by the measuring table inclination along the measurement path (Y axis). The best available tool for analysing angular error is a laser interferometer. Pitch and yaw measurements were made by making angular measurements at multiple points along the linear travel path in the Y axis as shown in Fig. 4. The angles of the measuring table were measured by using angular interferometer optics. The greatest Abbe errors were evaluated by analysing extreme angular deviations. For pitch (Fig. 4), the greatest angle differences along the measurement path of 500 mm was 7 ^m/m. The maximum offset of the laser reflector of 3 mm in the Z direction was estimated. Therefore the expected error interval for pitch is: 1 = 7 nm/mm • 3 mm = 21 nm (1) Standard uncertainty at supposed rectangular distribution for the Abbe error for pitch is: Fig. 4 Pitch and yaw measurements of the stage in Y direction 92 Advances in Production Engineering & Management 12(1) 2017 Automatic high resolution measurement set-up for calibrating precise line scales For yaw (Fig. 4), the greatest angle differences along the measurement path of 500 mm was 10 |j.m/m. The maximum offset of the laser reflector of 1 mm in the X direction was estimated. Therefore the expected error interval for the yaw is: / = 10 nm/mm • 1 mm = 10 nm (3) Standard uncertainty at supposed rectangular distribution for the Abbe error for the yaw is: 10 nm Uxy(